4.9.
Heating (Activation/Conditioning) without stop
Some applications require special getter pumps (HV) that work when the getter material is hot.
NEG POWER can be used to keep these pumps hot giving them the necessary power
continuously. To do this, t_hold =
is used (to set t_hold =
reduce t_hold down to 100 and
press down again). When t_hold =
the power supply is only stopped by pressing the stop
button.
4.10.
Custom: a Generic Pump for Customer Settings
NEG POWER software is preset to know the activation/conditioning parameter settings of the most
popular SAES getter pumps, like the GP SORB AC, the CapaciTorr, CapaciTorr HV and the
NEXTorr families (see the Summary Table of the NEG pumps controlled by the NEG POWER).
In the software there is also a Generic Pump named
Custom
.
If the used pump model isn’t included in the list, Custom pump can be selected and settings of
suitable voltage, current, power and temperature must be set for the specific pump model.
The default settings for Custom pump are the following. These values can and shall be changed.
In case of standard 700 W module:
Activation Mode
Conditioning Mode
NAME
Tmax Vmax Imax Pmax
Tmax Vmax Imax Pmax
Custom
500 °C
50 V
4.5 A
500 W
250 °C
25 V
2.5 A
250 W
In case of low power 150 W module:
Activation Mode
Conditioning Mode
NAME
Tmax Vmax Imax Pmax
Tmax Vmax Imax Pmax
Custom
500 °C
20.0 V
4.5 A
100 W
200 °C
5.0 V
2.5 A
15 W
ATTENTION: SAES Getters is not responsible for damages due to improper use of the
power supply