Table 2
Guidance and Manufacture’s Declaration – Electromagnetic Immunity
This equipment is intended for use in the electromagnetic environment specified
below. User should assure that it is used in such an environment.
Immunity test
IEC60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
±8kV contact
±2kV, ±4kV,
±8kV, ±15kV air
±8kV contact
±2kV, ±4kV,
±8kV, ±15kV
air
Floors should be wood, concrete
or ceramic tile. Humidity should
be at least 30% if it is synthetic
materials.
Electrical fast
transients/
bursts (EFT)
IEC 61000-
4-4
±2kV
100kHz
repetition
frequency
±2kV
100kHz
repetition
frequency
Main power quality should be
that of a typical commercial or
hospital environment.
Surges
IEC 61000-
4-5
±0.5kV, ±1kV
line-to-line
±0.5kV, ±1kV,
±2kV line-to-
ground
±0.5kV, ±1kV
line-to-line
±0.5kV, ±1kV,
±2kV line-to-
ground