P14 Display the code value
P15 High precision
P15 00
:
Off
P15 01
:
On
P16 Counting Speed
P16 00
:
Counting quickly, low accuracy
P16 01
:
Counting slowly, high accuracy
P17 ,18,19,20 Reserved function
Ⅸ
、
Capacity setting and internal calibration
operation
Press
,
open the scale and it show“CAL X”
,
press
to change methods.
1
:
One point correction
3
:
Three points correction
Press to choose unit
,
it shows“SP X”,press to
change
。
0
:
Unit is kg
1
:
Unit is g
Press to setting max capacity
,
it shows “FXXXXX”
,
press to change and to move.
Press to setting division
,
it shows“div X”press
to change.
1
:
The representative scale value is 1 jump
2
:
The representative scale value is 2 jump
5
:
The representative scale value is 5 jump
Press to setting decimal point
,
it shows“0.000”
press to change value.
0
:
In range as unit and the data has zero decimal point
0.0
:
In range as unit and the data has a decimal point
0.00
:
In range as unit and the data has two decimal point
0.000
:
In range as unit and the data has three decimal point
0.0000
:
In range as unit and the data has four decimal point
0.00000
:
In range as unit and the data has five decimal point
Press to go to next step.
Display shows“5-4-3-2-1-0”then into correction
,
The screen
16
17