background image

 

 
 

19 

8-Maintenance 

8.1 Considerations 

1)

 

Don’t  let  any  liquid  enter  the  equipment  in  order  to  prevent  electric 

shock. 

2)

 

It is prohibited to spray to clean/disinfect on the casing. 

3)

 

The  power  can  only  be  turned  on  when  interior  and  exterior  of  the 

equipment are completely dry. 

4)

 

Before  each  use,  please  check  air  tightness  of  the  rotor  cover  and  the 

microcentrifuge tube; the rotor cover and the microcentrifuge tube must 

be clean and undamaged. 

5)

 

High-temperature disinfection temperature should not exceed 121 °C for 

no  more  than  20  min.  After  disinfection,  please  apply  a  thin  layer  of 

journal grease onto the rotor cover screw thread. 

6)

 

Please  don’t  close  the  lid  when  storing  the  rotor  cover  or  the 

microcentrifuge tube. 

7)

 

It is prohibited to use any corrosive chemicals on the equipment and its 

fittings,  such  as  strong  and  weak  base,  strong  acid,  acetone, 

formaldehyde, halogenated hydroxide or phenol. 

8)

 

When polluted by corrosive chemicals, the equipment should be cleaned 

immediately with neutral cleaner. 

9)

 

Don’t  disinfect  the  equipment  with  UV,  β,  γ  ray  or  other  HERs.  Avoid 

storing the equipment in areas with strong ultraviolet radiation. 

8.2 Regular maintenance 

1)

 

Please  shut  down  the  microcentrifuge  and  unplug  the  power  supply 

before maintenance.   

2)

 

Regularly empty and clean the waste water tank. 

3)

 

Clean surfaces of the equipment and its fittings with contact to operators’ 

skin with gentle cleanser at least once every week. 

4)

 

If  the  equipment  is  used  to  process  samples  with  biological  hazards, 

please  regularly  remove  and  clean  the  rotor  and  disinfect  it  at  120  °C. 

Clean  the  rotor  with  ethyl  alcohol  or  ethyl  alcohol-containing 

disinfectant. 

Note:  If  rotors  with  non-detachable  sealing  ring,  which  cannot  be 

disinfected  for  more  than  50  times  and  should  be  replaced  when  they 

have been disinfected for 50 times.   

Summary of Contents for RWD-MB-001-A

Page 1: ...RWD MB 001 A 1 28...

Page 2: ...Microcentrifuge User Manual D...

Page 3: ...trademark or mark in this instruction manual are only for editing purposes without other purposes the rights belong to their respective rights owners Statement RWD reserves the right to modify the con...

Page 4: ...2 3 Operating restrictions 3 2 4 Operating safety 3 2 5 Safety symbols 3 3 Introduction of product structure 5 4 Product assembling 7 5 Operation instruction 8 5 1 Start up 8 5 2 Rotor Short 10 5 3 S...

Page 5: ...8 Maintenance 19 8 1 Considerations 19 8 2 Regular maintenance 19 8 3 Cleaning and disinfection 20 8 3 1 Clean and disinfect microcentrifuge interior 20 8 3 2 Clean and disinfect the rotor 20 8 4 Rep...

Page 6: ...and damages to the equipment If you have any question or suggestion regarding safety please contact RWD for after sales support 1 3 Comprehensive description The Centrifuge is a machine used to separa...

Page 7: ...ure 5 40 Humidity 15 75 non condensing Storage environment Temperature 20 60 Humidity 15 75 non condensing Air pressure 50 kPa 106 kPa Class of pollution 2 Operating voltage AC 220V 50 Hz rated voltag...

Page 8: ...e considered as violation of safety standards and intended use of the equipment RWD will not be responsible in any case your misuse of the equipment or failure to follow basic safety requirements 2 5...

Page 9: ...easures Wear protective suit such as particle protection mask gloves and protective shoe covers according to rules and regulations on processing of radioactive pollutants in your labs Dispose radioact...

Page 10: ...5 3 Introduction of product structure Fig 3 1 Fig 3 2 1 2 3 4 5 6 7 8 10 11 9...

Page 11: ...open microcentrifuge lid 6 Emergency switch If the centrifuge cannot open the lid manually after centrifugation loosen the emergency switch with a wrench to force open the lid 7 Fuse holder For instal...

Page 12: ...o the ambient temperature and ensure all power supplies meet equipment requirements 4 Connect the microcentrifuge to the power supply and press the power switch to turn it on Wait for the display scre...

Page 13: ...t centrifugation 4 After centrifugation open the equipment and take out the microcentrifuge tubes from the rotor 5 Leave the microcentrifuge open turn off the power and wait until the condensate insid...

Page 14: ...peed is the target speed during short run indicates that the set speed is the target speed during short run Sound Sleep Auto open continuous cooling Parameter setup icon Click to enter System Setup pa...

Page 15: ...his function is mainly used to lock homepage parameters to prevent changes 5 2 Rotor Short Click System Setup icon to enter Rotor and Short setup interface by default As shown in the figure the rotor...

Page 16: ...up Click or to enter the previous or next page in the system setup interface and click to enable other system functions such as Sleep Sound auto Auto open or continuous cooling or click to switch betw...

Page 17: ...for 8 hours and then cooling is automatically turned off to enter the sleep mode During continuous cooling the set temperature will be shown on the screen Whatever the set temperature the rotor cavity...

Page 18: ...ut in the microcentrifuge tube restore the rotor to its original position vertically and tighten the rotor nuts clockwise with rotor wrench until the rotor is firmly fastened nuts must be tightened fi...

Page 19: ...timing reaches the target centrifugation duration the operation stops automatically Click START STOP button to stop preset time run early before the target duration is reached 5 6 2 Short run Press SH...

Page 20: ...open the lid Device is unbalanced Check whether the samples and rotor are abnomal or else restart the device The motor is uneven with too large vibration 1 Check whether the sample is placed symmetric...

Page 21: ...C250 Check whether the input voltage is within the working range of AC180 AC250 Condenser hot Wait until the device is cool If still abnormal contact the after sales personnel Evaporator temperature i...

Page 22: ...l Brake resistance circuit fault Contact the after sales service Cooling fan failure Restart the device If still abnormal contact the after sales personnel Compressor fan stalls 1 Equipment 2 Restart...

Page 23: ...s during starting 1 The rotor is loaded asymmetrically 1 Shut down the microcentrifuge and load the rotor symmetrically 2 Restart the microcentrifuge Even though Short button is pressed the microcentr...

Page 24: ...such as strong and weak base strong acid acetone formaldehyde halogenated hydroxide or phenol 8 When polluted by corrosive chemicals the equipment should be cleaned immediately with neutral cleaner 9...

Page 25: ...rupture It is prohibited to apply grease on other parts of the equipment such as the lid lock the motor shaft or the rotor cone 6 Clean the motor shaft with a dry and dust free soft cloth 7 Check whet...

Page 26: ...ss should be recorded and kept after maintenance including time location maintenance procedures etc for future reference 8 6 Parts and materials If you find that some parts or materials should be repl...

Page 27: ...t Any damage caused by improper or over range use is not covered by the warranty If repair or replacement of parts is required relevant costs will be borne by the user When the equipment to be reworke...

Page 28: ...te 201 Dover DE 19904 Kent Delaware USA Add 19 20F Building 9A Vanke Cloud City III Liuxin 4 Street Nanshan District Shenzhen518000 Guangdong P R China Tel 001 858 900 6602 86 755 86111286 After sales...

Reviews: