background image

AUTO-ZERO SELECT

To auto zero the AMS810 remove the tubing from the pressure

ports (to remove the normal pressure source). Make sure that the

pressure ports are sheltered from any drafts, including the techni-

cian’s breath. Ideally the two ports should be connected together

with a short piece of tubing. Turning on switch 7 starts the auto-

zero process. Switches 1 through 6 may be in any allowed configu-

ration. The LED will blink at a fast rate and the output will be forced

to zero (4mA or 0V) while the auto-zero is being performed. When

the LED blinks at a slow rate, turn off switch 7 and reconnect the

pressure source.

SWITCH 8

Switch 8 is reserved for factory use, leave in the off posiiton.

TROUBLESHOOTING

PROBLEMS:

POSSIBLE SOLUTIONS:

LED does not light

- Check power connections for proper power

LED is blinking fast

- The unit may be performing an auto-zero. Wait 10 seconds and check again.

(

1

/

2

second on, 

1

/

2

second off)

- The dip switches are in an unsupported configuration, check the dip switch settings, both pressure

and output, and change them to ranges desired

Output stuck (high or low)

- Remove pressure from ports and perform auto-zero procedure

Output not tracking pressure properly

- Check dip switches for proper pressure range selection
- Check dip switches for proper output range selection

LED OPERATION

LED off

No Power

LED Blinking 

1

/

2

sec on, 

1

/

2

sec off

Auto-zero or dip switch in undefined configuration

LED on

Normal operation

LED Blinking 

1

/

2

sec on, 4 sec off

Auto-zero complete

Auto-Zero