10
11
ENGLISH
ENGLISH
Safety information
Do not perform maintenance or service on the device while it is in use. The patient
is an intended operator of the device. The device should be kept at least 30 cm away
from other wireless devices, such as WLAN units, microwave ovens etc to prevent
inaccurate readings. Swallowing batteries and/or battery fluid can be extremely
dangerous. Keep the batteries and the unit out of reach of children and disabled
persons. If you are allergic to plastic/rubber, please do not use this device. The
monitor can need up to 6 hours to cool down/warm up if stored at maximum/minimum
storage temperature until it is ready for it's intended use (20 °C ambient temperature).
Replace all batteries in the device when the display shows the battery warning. The
device is not intended for use on neonates, children or pregnant women. (Clinical
testing has not been conducted on neonates, children or pregnant women). Avoid
compression or restriction of the connection tubing during measurement, which may
cause inflation errors or harmful injury due to continuous cuff pressure. Avoid taking
measurements too often as it can cause injuries due to interference of the blood flow.
Consult your physician if you have any doubt regarding the following cases:
Application over a wound or inflammatory diseases. Application of the cuff on any
limb where intravascular access or therapy, or an arterioveous (A-V) shunt is present.
Application of the cuff on the arm on the side of a mastectomy. Simultaneous use
with other monitoring medical equipment on the same limb. Need to check the blood
circulation of the user. Avoid speaking or moving during a measurement. Stay still,
calm and rest for 5 minutes before taking a measurement.
Package Contents
Blood pressure monitor, cuff and storage box.
Proximity fields from RF wireless communications equipment
Test frequency
(MHz)
Band (MHz)
Immunity test levels
Professional healthcare
facility environment
385
380-390
Pulse modulation 18Hz, 27V/m
450
430-470
FM, ±5kHz deviation, 1kHz sine,
28V/m
710
704-787
Pulse modulation 217Hz, 9V/m
745
780
810
800-960
Pulse modulation 18Hz, 28V/m
870
930
1720
1700-1990
Pulse modulation 217Hz, 28V/m
1845
1970
2450
2400-2570
Pulse modulation 217Hz, 28V/m
5240
5100-5800
Pulse modulation 217Hz, 9V/m
5500
5785