Cobia Sense - User Manual v2015.1A

14

5.2

Battery Warnings

Warnings are shown when the battery is starting to run low (see below).
After the second warning, the system can suddenly shut down due to an
empty battery.

5.3

Communicating with a PC

The Cobia "Sense PC" model includes a user-friendly communication
protocol, allowing communication between the Cobia Sense and a
computer, including the Ocean version 2015 (or newer) software. The
protocol can also be integrated into other software. The protocol is
described in detail in the Cobia Reference Manual.

6

Troubleshooting

If you experience any problems with the Cobia, please go through the
following steps before contacting your local RTI distributor or RTI
Electronics:

A

. Check the RTI website for system updates (

www.rti.se)

.

B

. Go through the checklist below.

The Cobia does not start:

If the Cobia does not start, try the following: Connect the charger and
retry starting the Cobia. If it does start, then the battery was too low to
start. If this does not help, try resetting the Cobia as described below.

The external detector gives false readings:

Summary of Contents for Cobia Sense

Page 1: ...Cobia Sense User Manual English v2015 1A...

Page 2: ...errors or consequential damage that may result from the use or misinterpretation ofany information contained in this user manual Copyright 2011 2015 by RTI Electronics AB All rights reserved The cont...

Page 3: ...Time 9 2 2 Charging the Battery 9 10 3 Performing a Measurement 3 1 Measuring with an External Detector 10 3 2 Measuring with a Light Probe 11 11 4 Maintenance 4 1 Scheduled Calibration 11 4 2 Updatin...

Page 4: ...Cobia Sense User Manual v2015 1A 4 1 Product Overview TOP VIEW BOTTOM VIEW...

Page 5: ...tioning marker Not applicable for the Sense model 8 Measurement surface Not applicable for the Sense model 9 Status indicator Indicates functional status of the Cobia 10 External detector input Used t...

Page 6: ...ht side ofthe field displays an icon for external power supply and an indicator for the battery level 3 Asimple user guide is displayed at the bottom ofthe screen For each view the buttons that are fu...

Page 7: ...cling is in progress 2 Apulse indicator is shown below the play symbol This indicator is turned on when pulses are detected resulting in the display ofadditional pulse parameters List view In the List...

Page 8: ...urns the Cobia off When using an external detector or probe there will be an extra menu item Ext Detector that displays detector information and settings as described below Select External Detector Ca...

Page 9: ...ow button In List view Switches the display back to the Single view In Menu view switches the display back to the previous menu or screen 2 Getting Started 2 1 Before Starting the Cobia for the First...

Page 10: ...etector at a clinically relevant distance typically 50 100 cm from the source as described by the manual ofthe detector 4 Attach the external detector to the Cobia 5 Ifneeded select among the availabl...

Page 11: ...ly saved to the log function when using the shutter button For illuminance measurements the down arrow button has to be pressed for a value to be saved to the log Please note that there is no need to...

Page 12: ...and the Cobia 4 On the Cobia press the menu button and enter the System Info menu Make sure that the update file contains a newer system version than already present in the system 5 Ifso press and hol...

Page 13: ...a battery is charged On the front face ofthe Cobia there is an orange battery charging indicator marked The indicator is lit when the battery is actively charging It will gradually turn offwhen the ba...

Page 14: ...n also be integrated into other software The protocol is described in detail in the Cobia Reference Manual 6 Troubleshooting Ifyou experience any problems with the Cobia please go through the followin...

Page 15: ...nance within limitations stated below Ifinstalled according to accompanying documents the product is intended to be used together with all diagnostic X ray equipment except for therapeutical X ray sou...

Page 16: ...tudy ofthe manual studies ofthe built in help function in measurement software or on request in a short course ordered from the manufacturer The product is intended to be used inside X ray rooms ready...

Page 17: ...17 Cobia Sense User Manual v2015 1A NOTES...

Page 18: ...Cobia Sense User Manual v2015 1A 18 NOTES...

Page 19: ...19 Cobia Sense User Manual v2015 1A NOTES...

Page 20: ...ort support rti se Service service rti se Web www rti se Contact Information US Office RTI Electronics Inc 33 Jacksonville Road Bldg 1 Towaco NJ 07082 USA Phone 1 800 222 7537 Int 1 973 439 0242 Fax I...

Reviews: