Chamber Adapter User's Manual

2019-10-14/1.6A

15

Function and connection

Select type of ion chamber you want to add from the drop-down list.  Any chamber type can be used with the
Chamber Adapter that has the (PBCA) extension.  If you can't find your specific chamber, there are three
generic chamber types you can choose which are based on the chamber volume.

Enter the serial number of the chamber you wish to enter in the space below "Serial".

If you want to refer to your ion chamber by a different name than the one listed under "Type", you may
check the checkbox "Override default detector name" and type the name you want to call your chamber in the
space provided below the checkbox.

9.  When you click OK to continue, a new pop-up window is shown where you can select the calibration
details:

Select the calibration you want from the drop-down list (for our example, we will choose "Mo/Mo").  Enter the
calibration factor in the space provided.  You may select the units of measure and calibration date as well.

Ocean 2014 and the QABrowser assume that your chamber is calibrated at 20°C and 101,4 kPa.

Make necessary changes to the calibration factor if the calibration record for your chamber indicates different

temperature or pressure.

10.  Click OK to continue, the new calibration will be saved and all fields are cleared.  You may add more
calibrations if you want at this point.  

11.  To complete our example, just click Close.

12.  The final step to adding the new chamber is to store the information in the device.  To do this, select the
"Store to device" option on the screen. 

Summary of Contents for Chamber Adapter

Page 1: ...Chamber Adapter User s Manual English Version 1 6A RTI article number 4700 000784 Chamber Adapter Ion chamber adapter for Piranha...

Page 2: ...Chamber Adapter User s Manual 2019 10 14 1 6A...

Page 3: ...Chamber Adapter The Chamber Adapter makes it possible to use ion chambers with the Piranha and Cobia...

Page 4: ...f the product without prior permission from RTI Group AB RTI Electronics Inc 33 Jacksonville Road Building 1 Towaco NJ 07082 USA Phone 800 222 7537 Toll free Int 1 973 439 0242 Fax Int 1 973 439 0248...

Page 5: ...connect an ion chamber that is not intended for 300 V bias voltage to the Chamber Adapter The Chamber Adapter is only intended to be used with ion chambers that has a LEMO connector or with available...

Page 6: ...Chamber Adapter User s Manual 2019 10 14 1 6A...

Page 7: ...Function and connection 2 12 Function of the Chamber Adapter 2 1 12 Connecting the Chamber Adapter 2 2 13 Non RTI ion chambers 2 3 13 Adding a new ion chamber 2 4 17 Application example 3 18 Measurin...

Page 8: ...Chamber Adapter User s Manual 2019 10 14 1 6A...

Page 9: ...Chapter 1 Introduction...

Page 10: ...d for other applications For the purposes of this documentation the expression electrometer refers to the electrometer of the Piranha or Cobia The Chamber Adapter makes it possible to choose between s...

Page 11: ...Chapter 2 Function and connection...

Page 12: ...apter has a rechargeable battery The Piranha power supply not the Cobia power supply can be used to charge the Chamber Adapter To charge the Chamber Adapter 1 Turn off and disconnect the Chamber Adapt...

Page 13: ...TNT Adapter Cable 9730099 01 Malco to Lemo Adapter Cable You cannot connect ion chambers with a BNC Banana connector 2 4 Adding a new ion chamber The calibration factor for the preferred ion chamber...

Page 14: ...BCA is the chamber adapter identification information This information is locked at the factory during manufacturing and cannot be changed If you click on a detector to the left the available calibrat...

Page 15: ...x 9 When you click OK to continue a new pop up window is shown where you can select the calibration details Select the calibration you want from the drop down list for our example we will choose Mo Mo...

Page 16: ...nformation into the device Do not interrupt the program before this process is completed 13 Quit the Detector Manager when the new data is stored in the device As soon as you start Ocean 2014 or the Q...

Page 17: ...Chapter 3 Application example...

Page 18: ...pear The detector types that are available with the Chamber Adapter are identified by the text PBCA Piranha Bias Chamber Adapter If you are running Ocean 2014 in Keyboard mode all of them are availabl...

Page 19: ...tor Manager see Adding a new ion chamber section your own name is shown here instead 6 Select the ion chamber you want to use and click OK 7 The measurement is activated and you can start to make expo...

Page 20: ...on the Chamber Adapter 4 Activate the measurement CT Dose for example 5 The detector selection will appear Your particular detector selection picture may be different than those shown in this manual...

Page 21: ...nal button There are two drop down menus shown in the next window The upper menu shows the available probe types and when selected the lower part shows the serial number of the probe that is selected...

Page 22: ...Chamber Adapter User s Manual 2019 10 14 1 6A Application example 22...

Page 23: ...Chapter 4 Specifications...

Page 24: ...e Ripple 15 mV Power source Rechargeable battery Battery life 10 hours Current range 10 pA to 0 1 A 2 or 1 pA exp time 10 ms Air kerma rate DCT10 0 8 mGycm s 8 Gycm s 5 or 0 08 mGycm s exp time 10 ms...

Page 25: ...Chapter 5 FAQ...

Page 26: ...ang if it gets a very high signal for example when connecting or disconnecting the chamber I get the wrong time when measuring short times The Chamber Adapter is not suitable for time measurements esp...

Page 27: ...Applications CT 10 Mammography 10 B Battery 12 24 Bias 12 BNT 13 C calibration factor 13 Chargeing 12 D Detector Manager 13 I Internal detector 20 Ion chamber 13 calibration factor 13 connect 12 L Le...

Reviews: