OPERATION INSTRUCTION / EN
EN-8
09/16/11 Version No. 00
(c). Press “
↵
↵↵
↵
” key to enter the interval time setting mode and the three flashing numbers of
seconds.
(d). Press
or
key to set the desired second of interval time (1 to 255 seconds).
(e). Press “
↵
↵↵
↵
” key to store the auto datalogging interval time.
(f). Press “SET” key to exit this mode.
3. Offset setting:
You can adjust the meter readings to compensate for the errors of a specific thermocouple.
The allowable adjustment range is from +12.7 to -12.8 degree and regardless the
temperature units. You can store individual offsets for T1, T2, T3 and T4.
(a). Press “
” key to turn on the meter.
(b). Press “SET” key to enter the setting mode, then press
or
key until the display shows
“SEt OFSEt”.
(c). Press “
↵
↵↵
↵
” key to enter the offset setting mode, the “ OFFSET ” symbol is displayed.
(d). Press
⊳
or
key to select the desired T1, T2, T3 or T4.
(e). Press
or
key to set the desired offset values.
(f). Press “
↵
↵↵
↵
” key to store the offset value.
(g). Press “SET” key to exit this mode.
The temperature measurement plus the offset appears in the display. Remember to set the
offset to 0.0 when it is no longer needed. The “ OFFSET ” symbol will disappear when the
offset values all are 0.0.
4. Auto Power Off Time setting:
(a). Press “
” key to turn on the meter.
(b). Press “SET” key to enter the setting mode, then press
or
key until the display shows
“SEt SLEEP”.
(c). Press “
↵
↵↵
↵
” key to enter the auto power off time setting mode, the “SLEEP” symbol is
displayed.
(d). Press
or
key to choose the desired auto power off time. The choices are : 5, 15 and
30 minutes.
(e). Press “
↵
↵↵
↵
” key to store the choice.
(f). Press “SET” key to exit this mode.
5. Alarm High / Low Limit setting:
(a). Press “
” key to turn on the meter.
(b). Press “SET” key to enter the setting mode, then press
or
key until the display shows
“SEt ALArn”.
(c). Press “
↵
↵↵
↵
" key to enter the alarm high limit and low limit setting mode, the “ ALARM ”
symbol is displayed.
(d). Press
⊳
or
key to select the desired T1, T2, T3 or T4.
(e). Press “
↵
↵↵
↵
” key to enter the high limit value setting, the “
” symbol is displayed.