CH1/CH2 Independent Output
Instruction
CH1 and CH2 are working in independent mode and
insulated from the ground.
Output ratings
0~30V/0~3A (Max: 32V, 3.2A)
Operation steps
1. Make sure that parallel/series mode is off.
2. Connect load to the positive and negative terminals of CH1/CH2.
3. Set the voltage (current) value of CH1/CH2: Firstly, press V (or A )
to select the voltage (current) parameter needed to modify. Then
revolve the multi-function knob to set the wanted value (You
could press FINE to make accurate adjusting).
Coarse adjusting: 0.1V or 0.1A each step.
Fine adjusting: the least precision each step.
4. Open output: Press ON/OFF button to turn on the output, the
channel works in CC or CV mode and its indicator light gets
lit immediately.