15
Operating environmental conditions
Reference
Use
Storage
Temperature
+20
℃
~
+30
℃
0
℃
~
+50
℃
-30
℃
~
+70
℃
Relative Humidity
≤70%RH
10%
~
85%RH
10%
~
90%RH
(1) Dimensions and weight
:
69x26x165mm.500g
(2) Electrical safety to IEC 1010-1 Dual insulation
Installation category
Ⅲ
Degree of Pollution 2
Related voltage or max line-earth
:
6500V RMS CE
:
EN50081-1 and 50082-1
Operating procedure
Connect the probe to the oscilloscope with the BP-250 BNC TO BNC cable. Adjust the vertical
zero adjustment of the oscilloscope if necessary.
Select the attenuation rate and the vertical deviation of the oscilloscope in accordance with the
conversion table below.
Note: The power must be on
Attenuation Rate
X100
X10
Input Voltage(DC+AC Peak)
±800V
±80V
Attention: The real vertical deviation is equal to the attenuation rate multiplied by the
range of vertical deviation selected on the oscilloscope. It will be double in the case of a 50Ω
load.