Quick Start
RSDPB5000/RSDPB4000
High Voltage Differential Probe Series
Page 1: ...Quick Start RSDPB5000 RSDPB4000 High Voltage Differential Probe Series...
Page 2: ...2...
Page 3: ...e read this manual carefully before use Safety precautions 1 Be cautious of an electric shock 2 Pay attention to the maximum input voltage Please do not use in humid or in flammable and explosive envi...
Page 4: ...ng function this can also be closed manually with a USB power supply connector to make it easier and more flexible for use The probes are equipped with standard BNC input connectors they can be used w...
Page 5: ...ttenuation indicates different ranges such as RSDPB5150 500X indicate the maximum test voltage is 1500V 50X indicates the maximum test voltage is 1 5 0 V Oscilloscope attenuation factor should be set...
Page 6: ...scope oscilloscope input impedance should set to 1M Ifset to 50 the output attenuation is a half of the practical value 6 Power interface Standard USB type B interface supply power with standard USB a...
Page 7: ...262 Pincer Clips CK 281 CATIII 1000V Hook Clips CK 284 CATIII 1000V High Voltage Input differential lead CK 28 10A CATIII 1000V 10A CATIII 1000V 10A CATIII 1000V 10A CATIII 1000V Coaxial Output Line...
Page 8: ...Single ended to ground 5M Two inputs 10M Input Capacitance Single ended to ground 4pF Two inputs 2pF CMRR DC 80dB 80dB 80dB 100kHz 60dB 1MHz 50dB Noise Vrms 50X 50mV 500X 50mV 100X 200mV 1000X 1 2V P...
Page 9: ...lips CK 261 Approx 106 43 16 mm Pincer Clips CK 281 Approx 152 50 13 mm Hook Clips CK 284 Approx 121 23 23 mm Probe body dimensions 195 65 28 mm Probe body weight Approx 188 g Environment Characterist...
Page 10: ...probe body should be kept away from high voltage pulse circuits to reduce interference to the probe Test Mode Offset Setting Users may enter the test mode to adjust offset based on needs The probes m...
Page 11: ...DSO4000 Displays probe output Test bandwidth Standard signal Amplitude accuracy 0 75 rise time 3ns AC accuracy Generator or calibrator e g FLUKE WAVETEK 9100 Common mode rejection Digital millimeter A...
Page 12: ...ar 8 Repeat step 4 6 and calculate whether is within the accuracy ranges Model Attenuation Rate Signal source output voltage Probe expected output voltage RSDPB5150 50X 5V 100mV 2mV 500X 50V 100mV 2mV...
Page 13: ...o inputs of the digital millimeter 5 Enable signal source output respectively record voltage output values check with the following chart to calculate whether is within the ranges 6 Close the calibrat...
Page 14: ...scilloscopes When combined with the 50 load the attenuation will be 2 times as much RDPB4080 is recommend for use with our own manufactured PL 10 to expand the measuring with the electricity meter to...
Page 15: ...procedure Connect the probe to the oscilloscope with the BP 250 BNC TO BNC cable Adjust the vertical zero adjustment of the oscilloscope if necessary Select the attenuation rate and the vertical devia...
Page 16: ...h to clean Chemical must not be used to clean 3 Please put the probe in the package provided and put it in cool clean and dry places 4 Please put the probe in the package provided to prevent shock 5 D...