Milliohm Meter/ English
2016-9-10 / V001
EN-5
Precise Impedance measurement:
1.
Insert an RED testing lead into “HI” socket and another RED testing lead into
the adjacent “RX” socket; insert a BLACK testing lead into “LO” socket and
another BLACK testing lead into the adjacent “RX” socket.
2.
Select the function switch to 200 m
Ω
position.
3.
Short circuit the two testing clips, adjust “ZERO ADJUST KNOB” to a ”000”
reading value on the LCD.
4.
Whenever a selected range of 200 m
Ω
, 2
Ω
, 20
Ω
, 200
Ω
to be measured the
readjustment of
“ZERO ADJST” shall be deeply necessary.
5.
To measure an unknown resistor RX then by following the two point type of
measurement per shown in figure 3.
FIG (3)
6.
Measure the resistance on PCB by using the two-point type of measurement
per shown in figure
FIG (4)
7.
The measured value is shown on LCD