S1
&
S2 & S3 / EN
25
(16) Capacitance
Range
OL Reading
Resolution
Accuracy
1.000µF
1.100µF
0.001µF
±
(1.9% + 5D)
10.00µF
11.00µF
0.01µF
±
(1.9% + 2D)
100.0µF
110.0µF
0.1µF
1.000mF
1.100mF
0.001mF
10.00mF
11.00mF
0.01mF
Overload Protection :
AC/DC 1000V
(17) Frequency
Range
OL Reading
Resolution
Accuracy
100.00Hz
100.00Hz
0.01Hz
±
(0.1% + 2D)
1000.0Hz
1000.0Hz
0.1Hz
10.000kHz
10.000kHz
0.001kHz
100.00kHz
100.00kHz
0.01kHz
Minimum Sensitivity (Ampere) :
>0.6A
(18) VoltSeek
Voltage Range of High Sensitivity :
80V ~ 1000V (At the top edge of the meter)
Voltage Range of Low Sensitivity :
160V ~ 1000V (At the top edge of the meter)
(19) HFR (High Frequency Rejection)
Available for AC only.
Add ± 4% to specified accuracy of each function and each range
for 45Hz to 200Hz.
Accuracy is unspecified for > 200Hz.
Cut-off Frequency (-3dB) :
800Hz
Minimum Sensitivity (Voltage) :
Range
Accuracy
1Hz – 10kHz
>5V
10kHz – 50kHz
>20V
50kHz – 100kHz
unspecified