1. Does the vacuum pump work?
▪ Check the vacuum pressure by using the option
Check flow sensor
in menu
.
2. Rinse vacuum valve not working, check functioning and if needed, replace, only allowed
for a trained
technician
.
3. Waste separator leaking, remove, disassemble and check separator and reassemble.
4. Rinse nozzle not aligned,
▪ Re-alignment for the rinse nozzle needed, only allowed
for a trained technician
.
15.2 - Liquid level sensor not sensing
1. Liquid in the container is not detected. This occurs sometimes with the
De-ionized water
bottle and is
caused by a very low conductivity.
2. Add one or two drops of
Saline
to the
De-ionized water
to increase the conductivity.
15.3 - Starrsed RS stalls
When the Starrsed RS is not working, check the main fuse. The main fuse can be found at the main input
socket from the ESR Unit.