13.3.2 - Replace bacterial filters
Detailed instructions of this procedure can be found in
As part of the Cleaning waste separator procedure the bacterial Hepa filter
QWLV040002
is replaced with a
new one.
Exchange bacterial filter
QWLV040001
on the waste bottle assembly.
13.3.3 - Fill-nozzle O-ring replacement
As the fill nozzle O-ring (
QWLV050004
) ages, it looses its flexibility and air-bubbles may occur in the
Westergren pipettes, the O-ring needs to be replaced.
Symptoms for a bad fill-nozzle O-ring
After the aspiration, the Westergren pipette has a zebra pattern (air- blood- air -blood, nicely divided in the
column.)
Vacuum stabilisation errors may occur.
13.3.4 - Fill and clean procedure
Note
: Each pipette on the pipette belt will be filled with Starrsed Cleaning Agent, after one hour the first
pipette is washed and dried.
Fill and clean with adapter:
Cleaning agent preparation Starrsed RS: Fill and clean:
This cycle takes about 90 minutes.
1. Fill the clean adapter VERA119002 with hot
de-ionized water (80 °C) till the first mark in the
adapter. (180 ml).
2. Add cleaning agent (QRR 010905) till the
second marker in the adapter. (18 ml).
3. Place the two caps on the adapter and mix
well.
4. Put the adapter in the Input-pool.
5. Select
Maintenance -
Prime/Clean,
button
Fill and
clean
.
General maintenance
134
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...