22
PX5 PAPR
®
Speed Settings: Cycle through the 3 speed
settings by pressing the Power Button. Each
short button press will advance through these
speed settings.
FAN SPEED
FIGURE 8.3
LOW AIRFLOW ALARM
A continuous alarm sounds when the airflow to the hood falls
below the minimum requirement of 6cfm (170slpm) for more
than 5-10 seconds in duration.
Leave the work area to check the filter for blockage or
anything obstructing the air path to the head-top. If the filter
is blocked due to contaminants, replace the filter with a new
one.
LOW BATTERY CHARGE
ALARM
A pulsing alarm sounds when approximately 10-15 minutes
of charge is left in the battery. Leave the work area. Power
the unit off and remove the battery and place on the charger,
if you have a second battery you can insert it into the device
to continue operation. The pulsing alarm will also sound if
the battery is overheating. If the unit is operated outside the
operating temperature range for 10 minutes the battery alarm
will activate.
Alarm volume is 85 dBA at 4” (104mm).
ALARMS
TABLE 1.1
Filter blockage is indicated on the User
Interface Panel by the multi-colored arrow
under the power button. Green = full airflow,
Yellow = mild blockage, reduced airflow, Red
= blocked airflow. When the Red arrow is lit,
safely exit the work area and check for any
blockages in the air path. If blockage is not
resolved, follow the filter change instructions.
FILTER BLOCKAGE
FIGURE 8.4
RESPIRATOR SETUP AND CARE
CONTINUED