IN-E-HyFlex-V4_10
Rotronic AG
Bassersdorf, Switzerland
Document code
Unit
Instruction Manual
Document Type
HygroFlex transmitter version 4: instruction
manual
Document title
Page
41
of 61
© 2006; Rotronic AG IN-E-HyFlex-V2_10.doc
Procedure
The Remote Control mode is automatically activated by connecting any
probe input of the HygroPalm to the Test port of the transmitter
HygroPalm 1: remove the probe and replace it with cable #
AC1621
HygroPalm 2 and 3: connect cable
#AC1620
to the second probe input
As soon as the two instruments are connected, the HygroPalm detects the
transmitter and this is indicated by a flashing star on each side of the LC
display. At that time, the display of the HygroPalm shows data from the
remote unit and can no longer be used to access local data from the
HygroPalm. Similarly, the function keys of the HygroPalm control the
transmitter and not the HygroPalm itself. To exit the Remote Control mode,
simply disconnect the two instruments.
Notes:
Using the Remote Control mode does not disturb the output signals (analog
or digital) of the transmitter. The outputs keep being normally updated. The
HygroPalm cannot be turned off while in the Remote Control Mode.
9.1 Display the Probe Measurements
As soon as the Remote Control mode is initiated, the HygroPalm displays the measurements from
the probe(s) connected to the transmitter. When two probes are connected to the transmitter, each
probe can be displayed with the UP or DOWN key on the HygroPalm.
9.2 Functions (except ADJUST M.PT and ADJUST 1PT)
Press the MENU key of the HygroPalm to access the functions of the transmitter and use the UP
and DOWN keys to navigate the function menu. Selections are confirmed by pressing the ENTER
key.
Access to the functions depends on the model of HygroPalm. The functions CALCULATE and
DISPLAY are generally not accessible in the Remote Control mode. For more details see
HygroFlex Function Menu
9.3 Functions ADJUST M.PT and ADJUST 1PT
The functions ADJUST M.PT
1)
and ADJUST 1PT
1)
can be accessed only with the HygroPalm 3. In
the remote mode, both the display and the keyboard of the HygroPalm act like if they were the
display and keyboard of the transmitter. The definition of these functions and the procedure to be
followed are the same as already described under HygroFlex Function Menu.
1) applies only to the ROTRONIC HygroClip digital probes