7.5 Probe
adjustment
Please consult the HW4 on-line instruction manual.
8.
Baud rate compatibility requirements
Depending on the model, devices from ROTRONIC use one of the following Baud rates as the
factory default:
•
57600 bps: HygroLog NT data logger and docking stations
•
19200 bps: HygroPalm and HygroLab indicators, HygroFlex and M33 transmitters,
HygroClip Alarm programmable alarm card, HygroClip DI interface, HygroStat MB Thermo-
Hygrostat
IMPORTANT:
•
All devices connected to a RS-485 multi-drop network (master and slaves) must use the
same Baud rate
•
In the case of a device with an Ethernet (TCP/IP) module, the Baud rate of the module
must be the same as the Baud rate of the device
Devices and internal Ethernet modules with mismatched Baud rate will not communicate
together.
•
The Baud rate configuration of a ROTRONIC device can be changed with the
Device
Manager
function (Digital Interface tab) available from within the HW4 software.
•
The Baud rate of a Digi Module can be changed only from the module web interface. See
separate document
IN-E-TCPIP-Conf_10.doc
(also available in PDF format).