1400268-9S_D14.754.19.03_2019-07-20 QuattroPowerSupportKid Eng
Page
7
from
12
•
The support handles is resistant to the active agents in the hand and surface disinfectants certified by the DGHM
(Disinfection Commission of the German Society for Hygiene and Microbiology) with the exception of phenol
ether, pyridine derivative and thiocyanate.
•
Only alcoholic disinfection without re-fattening component may be used for the rubber discs.
5.2 Maintenance
(service):
•
If the handle has been attached to the wall for a prolonged period of time, the rubber discs will overstretch. We
therefore recommend that you remove the handle from the wall approx. every 14 days and leave it overnight on a
flat surface with the rubber washers relaxed, then reattach it the next morning.
•
Examine the rubber pads for possible functional deterioration because of reduced elasticity due to age, wear,
exposure to sunlight, uneven storage, scoring, etc.
If the rubber pads are damaged and therefore the adhesive strength is limited, the complete suction heads (spare
part no. 14002S) are to be replaced or the rubber pads (spare part no. 1402721) must be exchanged by trained
specialist staff.
•
Keep the vacuum safety indicator always free from dirt!!
•
Check the handle regularly for damage and do not use the handle under any circumstances if it is
damaged.
6. Recognizing and remedying malfunctions
Malfunctions
Possible causes of
malfunctions/test
Corrective measures
With new handle: safety
indicator pops up after a
short time
Lime or other dirt on the surface.
Clean the surface thoroughly.
The surface is not non-porous even if it
looks smooth, or it is damaged (e.g. hairline
crack in the tile).
Possible test procedure: Attach the handle
to a glass pane or mirror and pay attention
to the safety indicator.
If now the safety indicator does not move
any more, the fault is in the substrate.
If possible, attach the handle at another location
where the underground is non-porous.
If this is not possible, self-adhesive stainless-steel
plates (order no. 1461418) to which the handles can
be attached can be glued on. See accessories.
With older handle: safety
indicator pops up after a
short time
Rubber discs scratched or damaged.
If there is damage, the rubber discs (order no.
1402721) can be replaced by a specialist dealer or
the complete suction head (order no. 14002S) can
be replaced.
Indicator rocker is still
slightly raised after
mounting
1. Toggle lever not firmly pressed down
before turning over!
Attach the handle again while beforehand pressing
the rocker arm firmly down. (Page 8 fig. 3 arrow 2)
2. The rubber pad is overstretched.
This happens when the handle has
been attached to the wall for a long
time.
Store the handle overnight on a flat surface with the
rocker arm pointing upwards. Attach again the next
day while beforehand pressing the rocker arm firmly
down. If the mounting rocker still protrudes after
pressing it in, the rubber discs (order no. 1402721)
must be replaced by a specialist dealer or the
complete suction head (order no. 14002S) must be
replaced.
The suction cups cannot
be pressed flat against
the wall
The built-in safety brakes prevent the
suction heads from rotating.
With sufficient force, the suction heads can be
adjusted correctly. If this is not possible for you,
please ask a specialist to help you.
The quick lock has fallen
out.
The quick-release latches (page 9 fig. 8)
are pressed in without the red additional
part.
To unlock another object taken which was
too long.
The quick-release locks can be quickly reinstalled by
a specialist using a tool.
If these are lost, individual parts can also be
reordered.
7. Disposal:
The product packaging should be recycled. The metal parts can be recycled as scrap metal. Disposal must be carried
out according to the respective national statutory regulations. Please ask your local or communal authority about local
disposal facilities.
These ROTH mobile grab handles may not be exported to the U.S.
Subject to technical modification.