9
Figures in this section provide the guidelines for wiring
the 396P-01 sensor to various Analyzer/Transmitter
instruments.
To determine which wiring guideline to use, locate the
model number of the sensor to be installed.
1.
If the cable needs to be extended, use a high
quality eleven conductor double shielded instru-
ment cable available from Rosemount Analytical.
Refer to Table 3-1 for the appropriate junction box
to use and the corresponding wiring details.
NOTE
If the cable is too long, loop up the excess
cable. If the cable has to be shortened, cut
and terminate each conductor neatly and
make sure that the overall (outermost)
drain wire is not shorted out with either
of the two inner drain wires (shields).
2.
Signal cable should be run in a dedicated conduit
(preferably an earth grounded metallic conduit)
and should be kept away from AC power lines.
For your convenience, a wire nut kit is furnished
(in a plastic bag wrapped around the cable).
NOTE
For maximum EMI/RFI protection when
wiring from the sensor to the junction box,
the outer braid of the sensor should be
connected to the outer braided shield of
the extension cable. The outer braid of the
extension cable to the instrument must be
terminated at earth ground or by using an
appropriate metal cable gland fitting that
provides a secure connection to the instru-
ment cable.
Preamplifier Location
Analyzer/Transmitter
Sensor
Integral to Integral to Remote
Interconnect 1054, 1054A, 54, 81,
1055
1181
1050
1003
SCL or
Refer to
Model
Sensor
Analyzer
Junction
(unprepped 1054B,
2054
3081,
1060
1023 SoluCube
Figure
Options
Box (PN)
PN)
2081
4081
#
396P-01-()-50
✓
✓
✓
3-2
396P-01-()-50
✓
Ext. box
9200254
✓
✓
3-2
(22719-02)
396P-01-()-54
✓
✓
3-3
396P-01-()-54
✓
Ext. box
9200254
✓
3-3
(22719-02)
396P-01-()-55
✓
✓
3-5
396P-01-()-55
✓
Ext. box
9200273
✓
3-4
(23550-00)
396P-01-()-55
✓
9200273
✓
3-6
or 23646-01
TABLE 3-1. Wiring Matrix Guide
MODEL 396P and 396PVP
SECTION 3.0
WIRING - MODEL 396P
SECTION 3.0
WIRING — MODEL 396P-01