Appendix B
Product Certifications
Rosemount 3144P With HART / 4–20 mA . . . . . . . . . . . . . page B-1
Rosemount 3144P With F
OUNDATION
fieldbus . . . . . . . . . page B-9
Installation Drawings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page B-17
ROSEMOUNT 3144P WITH HART / 4–20 mA
Approved Manufacturing
Locations
Rosemount Inc. – Chanhassen, Minnesota, USA
Rosemount Temperature GmbH – Germany
Emerson Process Management Asia Pacific – Singapore
European Union
Directive Information
The most recent revision of the European Union Declaration of Conformity
can be found at www.emersonprocess.com.
ATEX Directive (94/9/EC)
Rosemount Inc. complies with the ATEX Directive.
Electro Magnetic Compatibility (EMC) (2004/108/EC)
EN 61326-2-3:2006 and EN 61326-1:2006
Hazardous Locations
Installations
North American Certifications
Factory Mutual (FM) Approvals
E5 FM Explosion-proof, Dust Ignition-proof and Non-Incendive
Certificate Number: 3012752
Explosion-proof for Class I, Division 1, Groups A, B, C, D.
Dust Ignition-Proof for use in Class II/III, Division 1, Groups E, F, and G.
Temperature Code: T5 (T
amb
= – 50 to 85 °C)
Explosion-proof and Dust Ignition-proof when installed in accordance
with Rosemount drawing 03144-0320. Indoor and outdoor use. Type 4X.
NOTE
For Group A, seal all conduits within 18 inches of enclosure; otherwise,
conduit seal not required for compliance with NEC 501-15(A)(1).
Non-incendive for use in Class I, Division 2, Groups A, B, C, and D.
Suitable for use in Class II/III, Division 2, Groups F and G.
Temperature codes: T5 (T
amb
= – 60 to 85 °C)
T6 (T
amb
= – 60 to 60 °C)
Non-incendive when installed in accordance with Rosemount drawing
03144-0321.