10. Technical data:
Coding: Zone 0/1
Ex d IIC T6 (Ta = -40°C to + 75°C)
Ex d IIC T4 (Ta = -40°C to +125°C)
Ex d IIC T3 (Ta = -40°C to +150°C)
Ex tD A21 T85°C (Tamb -40°C to +75°C) IP6X
Ex tD A21 T135°C (Tamb -40°C to +125°C) IP6X
Ex tD A21 T160°C (Tamb -40°C to +150°C) IP6X
Ta = the higher of the process or ambient temperature.
For electrical details and pressure ratings, refer to Section 2: Installation and
Appendix A: Reference Data.
11. Label and certification plate details.
Typical arrangement shown. Sensor specific details omitted.
12. Cable selection.
It is the responsibility of the user to ensure that suitably temperature rated cable is used.
The table below is a guide to selection:
T Class
Cable Temperature Rating
T6
85°C
T5
100°C
T4
135°C
T3
160°C
WARNING
ARNING
DO NOT OPEN WHEN AN EXPL
O NOT OPEN WHEN AN EXPLOSIVE
OSIVE
ATMOSPHERE MA
TMOSPHERE MAY BE PRESENT
Y BE PRESENT
REFER TO
REFER TO
INSTRUCTION
INSTRUCTION
MANUAL
MANUAL
FL
FLAMEPROOF VIBRA
AMEPROOF VIBRA TING FORK LEVEL SWIT
TING FORK LEVEL SWIT CH
CH
ROSEMOUNT 2120
ROSEMOUNT 2120
R
US
E
E
L
B
S
A
UI
C
TA
D
E
B
T
L
A
Y
R
E
T
E
UR
M R
E AT
P
u
mo
s
n
e
.
o
t c
r.
w
om
w
w
Made in UK
Made in UK
Made in UK
ENCL. IP66/67
ENCL. IP66/67
CONDUIT
CONDUIT
ENTRIES M20
ENTRIES M20
ZONE 0/1 IECEx SIR 06.0051
Manufactured by Mobrey Lt d, Slough, SL1 4UE
REFER
TO
MANUAL
Ex d IIC T6 Ex tD A21 T85°C (-40°C Ta +75°C)
Ex d IIC T4 Ex tD A21 T135°C (-40°C Ta +125°C)
Ex d IIC T3 Ex tD A21 T160°C (-40°C Ta +150°C)
≤
≤
≤
≤
≤
≤
Model No.
Summary of Contents for 2120
Page 2: ......
Page 4: ...Reference Manual 00809 0100 4030 Rev DA June 2007 Rosemount 2120 ...
Page 8: ...Reference Manual 00809 0100 4030 Rev DA June 2007 Rosemount 2120 TOC 4 ...
Page 68: ...Reference Manual 00809 0100 4030 Rev DA June 2007 Rosemount 2120 B 20 ...
Page 69: ......
Page 70: ......
Page 71: ......