30
MODEL 1054B pH/ORP
SECTION 4.0
CONFIGURATION
4.10 ALARM SETPOINT.
The alarm setpoints should
be adjusted after completing the configuration proce-
dure as outlined in
1.
key to ensure that the analyzer is
2.
or
SP2
will show briefly, followed by the Alarm 1 or
Alarm 2 setpoint.
If the alarm is set to OFF or FAULT
(Alarm 2 only), the analyzer will display
NOTE
Alarm logic may be changed from normal-
ly open (N.O.) to normally closed (N.C.) by
cutting circuits on the power supply PCB
(W-5, W-7, W-9) and adding jumpers (W-
4, W-6, W-8).
3.
SELECT
to adjust the value. The display will
acknowledge briefly with
AdJ
followed by the
Numeric Display with digit flashing.
4.
SCROLL
(
é
) and
SHIFT
(
ç
) to display the
desired value.
5.
ENTER
value into memory.
6.
Repeat Steps 2 to 5 for the second setpoint.
ZERO
ALARM
1
F.S.
ALARM
2
ACCESS
é
AdJ
SP1
/
2
ç
SELECT
ç
SELECT
ENTER
Press
Once
Press
Once
Displays
Briefly
Displays
Briefly
Numeric
Display
Change to
desired
value
Press
Once
Numeric
Display
of
Setpoint
FIGURE 4-4. Alarm Setpoint