Page 1: ...Ropox Nursing Table Medi 2 User Manual Mounting Instructions Keep this manual with the product at all times TF 200 01 0052_eng Date 28 09 2022 ...

Page 2: ...ning of water supply drain and power 10 4 2 List of components 11 4 3 Installation of product 11 4 3 1 Dimensions 12 4 3 2 Mounting instructions 12 4 4 Wire diagram 19 4 5 Residual risks 20 5 Performance test 21 6 Safety instructions 21 7 Troubleshooting 22 8 Service and maintenance 22 8 1 Cleaning frame 22 8 2 Cleaning tabletop 22 8 3 Maintenance 23 8 4 Service schedule operation and maintenance ...

Page 3: ...understood before use Always keep this manual in close proximity of the product The use installation and service of this product must be in compliance with this manual to avoid accidents and serious personal injury Never use or handle this product in other ways specified in this manual as it can result in personal safety hazards and or cause damage to the product Persons installing and or using th...

Page 4: ...or have become damaged before being repaired or replaced If the control box makes unusual noise or smells switch off the mains voltage immediately Take care that the cables are not damaged The products must only be used in an environment that corresponds to their IP protection The information in this manual is based on correct installation in accordance with installation instructions for this prod...

Page 5: ...ction Stock number Serial number Operating interval of electrically operated equipment The use of electrical height adjustable equipment can run at maximum 2 minutes followed by an 18 minute break Type B applied part Consult manual for important safety related information warnings and safety precautions Class II double isolated electrical components Do not dispose as unsorted municipal waste Produ...

Page 6: ...0 14684 50 50694 50 50702 50 50690 50 50687 50 50697 50 50696 Mattress for nursing table 90cm Mattress for nursing table 130cm Adapter frame for nursing table Medi 2 20cm from wall Side covering white laminate 82x105cm Side covering white laminate 61x105cm Side bed guard for nursing table 130cm Side bed guard for nursing table 90cm Mixer tap Touchless Mixer tap with extension shower head 2 step la...

Page 7: ...sential performance The product does not have essential performance as the patient and user do not experience any risk in case of loss of product function performance 3 6 Nonclinical functions Height adjusting offers better ergonomics to the user 3 7 Clinical functions The product does not have any clinical functions 3 8 Product dimensions 50 50676 130cm Note All dimensions in mm ...

Page 8: ...8 TF 200 01 0052_eng 50 50678 90cm Note All dimensions in mm ...

Page 9: ...d Hand control with spiral cable option Warning The product should always be able to travel the full range of actuation without colliding with objects Failing to do so will compromise the product stability and introduce risk of squeezing body parts Special care should be taken in ensuring that no children or adults are sited under the product as severe injury from entrapment may occur ...

Page 10: ... screw Power Recommended positioning of power supply point in wall can be seen below Water The drain is Ø40 The product comes with an Ø40 Ø32 drain reducer that can be used if necessary We recommend positioning of water and drain installations as illustrated All hoses for water and drain installations must be flexible to ensure unhindered movement within the height adjustment range of 60 100 cm 4 ...

Page 11: ...ion of product Warning Before use always make sure that mains cable and hand control cable will be able to travel freely without the risk of squeezing or disconnecting the product as result of the movement Caution When connected to the mains the product must be placed in a manner according to power outlets so that the operator will be able to disconnect the product during use in case of an emergen...

Page 12: ...ame and drill the holes The nursing table is supplied together with a set 8 pcs of screws and rawlplugs which is used to mount the unit in a concrete wall The installer must always examine the material condition and strength of the wall and use appropriate screws and rawlplugs Caution Ensure no dust or other obstacles are interfering with the rollers chain sprocket or gas springs This can cause ma...

Page 13: ... been determined where the unit should be mounted in the room the unit is mounted on the wall Make sure the unit is level Also keep in mind where the power will be placed If the support legs have been purchased they must be mounted before the lifting unit is mounted on the wall See separate manual ...

Page 14: ...14 TF 200 01 0052_eng 2 Re mount the outer cover ...

Page 15: ...the tabletop has been mounted measure between the tabletop and the wall There must be 4mm of distance at most If it is more than 4mm there s a risk of entrapment squeezing Adjust the tabletop by pulling up on it it s mounted with Velcro When the top is free position it such that after re attaching it there is 4mm of distance or less ...

Page 16: ...16 TF 200 01 0052_eng 4 Connect the safety stop cable as shown 5 The water and drain can now be mounted Drain hoses and water hoses come with the version with a sink ...

Page 17: ...17 TF 200 01 0052_eng ...

Page 18: ...18 TF 200 01 0052_eng 6 The nursing table is now installed If the tabletop unit is still not level it s possible to adjust it This is done by tightening loosening the screws in the sides Or ...

Page 19: ...e is turned in the correct direction before plugging it into the control box The cable must be pressed into the plug until the O ring is no longer visible The motor cable is connected to input 1 in the control box Input 2 3 and 4 is unused and is mounted with sealing plugs The hand control is connected to input HB The mains cable is connected to the side of the control box 4 1 3 5 2 1 Control box ...

Page 20: ... nursing table keep away from these while the table is being actuated Warning Be aware of loose cables and hoses Do not let children play near these as they constitute a strangulation hazard if the cable is twisted around the neck Warning In case of incorrect mounting of plumbing parts a slipping hazard may occur as water will spill on the floor increasing the risk of falling ...

Page 21: ...rd 10 Press DOWN on the control switch and let the frame move 2 5 cm downwards then activate the safety stop The frame must stop downwards movement If all these tests are ok the nursing table is ready for use See section Safety instructions 6 Safety instructions Never leave a child unattended on the nursing table without supervision not even for a short period of time The nursing table may only be...

Page 22: ...thing is stuck and preventing the height adjustment Water doesn t drain from sink fast enough Check the water siphon for clogging If it s clogged clean it Check to see if the drain hose has been placed too high on the wall If it has it might create a water lock with the nursing table in certain positions Refer back to Installation Interfaces 8 Service and maintenance 8 1 Cleaning frame Simply dry ...

Page 23: ...ned Check that the frame moves freely from bottom to top position without problems Check that water and waste pipes are tight and undamaged Check that all cables have been fitted correctly and are undamaged After each inspection the service schedule shall be filled in Note Ropox will make parts list and circuit diagrams available for maintenance and repair beyond what is described in this manual O...

Page 24: ...ate Signature Remarks Service and maintenance Serial No Date Signature Remarks Service and maintenance Serial No Date Signature Remarks Service and maintenance Serial No Date Signature Remarks Service and maintenance Serial No Date Signature Remarks Service and maintenance Serial No Date Signature Remarks Service and maintenance Serial No Date Signature Remarks ...

Page 25: ... may be void 10 Accessories For Ropox Nursing Table Maxi 2 the following products can be purchased as accessories Do Note some accessories must be ordered together with the product This will be in the description Item number Description Image 50 50680 Mattress for Nursing Table 90cm 88 x 70 x 4 cm 50 50681 Mattress for Nursing Table 130cm 88 x 75 x 4 cm 50 50689 Adapter frame for Nursing Table Med...

Page 26: ...ting on floor wall 50 50688 Side Bed Guard for Medi2 Maxi2 130 cm 82 x 24 cm Mounted on worktop Not for retrofitting 50 50699 Side Bed Guard for Medi2 Maxi2 90 cm 82 x 24 cm Mounted on worktop Not for retrofitting Image shows two side bed guards mounted 50 14684 Mixer tab Non touch monobloc tap Iqua 50 50694 Mixer tab with extension shower head 50 50702 2 step ladder White ...

Page 27: ...ne by waste disposal professionals Part Disposal recommendation Lifting unit frame Steel Mixer tap Steel Cover Plastic Control box Electronic Actuator Electronic 50 50690 Extra hand control with spiral cable For Maxi 2 and Medi 2 with safety stop 50 50687 Support legs for Medi 2 90 130cm 50 50697 Ropox Basket unit without wash basin 50 50696 Ropox Basket unit with wash basin ...

Page 28: ...erating normally 12 3 Cables Cable component Length Power cable 3 2m Hand control spiral cable Approx 5m extended Warning Use of accessories transducers and cables other than those specified or provided by the manufacturer of this could result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation 12 4 RF portable equipment W...