background image

TF 200.06.0003_ENG 

17 

 

6. 

Maintenance 

 

 

 

 

Notification!

 

Failing to comply with periodic maintenance may result of degrading of product function and 
safety. 

 

 

 

 

 

 
Notification!

 

 

Ropox will make parts list and circuit diagrams available for maintenance and repair beyond 
what is described in this manual. Repair beyond what is described in this manual may only be 
done by Ropox designated service personnel.  

6.1 Periodic maintenance of bed guards 

If the product is used in a humid or dry atmosphere the locking pin may be prone to failure from 
lubricants drying. Besides checking the function of the locking mechanism before every use, a periodic 
lubrication of the locking pin every 6 month is recommended. 

6.2 Periodic charging of battery 

The battery must be periodically charged minimum every 6 months to sustain capacity. If the product is 
left unused or the battery is disconnected this may result degrading of the battery capacity. Charging the 
battery minimum every 6 months will mitigate this. 

6.3 Replacement of battery 

 

 
 
 

 
Notification! 
 

The battery is to be replaced after 4 years at the latest.  

6.4 Fault finding 

6.5 Calibration of lifting columns 

Lifting columns are aligned and calibrated from the factory. In case they need calibration to run 
simultaneously use the 

“Reset”

 procedure. 

Problem 

Check the following 

Mobilio does not move up or down when the hand-
control switch is pressed 

1.  The product is connected to 230V and the 

control box LED indicator lights green 

2.  Motor cables are connected to control box 
3.  The control switch has been connected with 

the control unit 

4.  If the product is used with battery, check 

whether it is charged. 

5.  The motors are overloaded (note: 2 min. 

ON, 18 min OFF). 

Control switch defective 

1.  Order a new control switch 

Mobilio is not horizontal / level 

”Reset”  

Press and hold UP (     ) and DOWN (     ) buttons. 
Keep the buttons depressed until the control unit 
has beeped approx. ten times. After that, the 
product may be adjusted to the bottom position for 
resetting. 
 

Summary of Contents for 40-31406-1

Page 1: ...Ropox Mobilio Showerbed User manual This manual should always be in close proximity of the product TF 200 06 0003_ENG Date 25 07 2022 ...

Page 2: ...fety in use 13 4 2 5 Ambient temperature 13 4 3 Installation of battery 14 4 3 1 Mounting the battery 14 4 3 2 Connecting the battery 14 4 3 3 Charging the battery 15 4 3 4 Battery feedback indicators 15 4 4 Electric block diagram 15 5 Cleaning 16 6 Maintenance 17 6 1 Periodic maintenance of bed guards 17 6 2 Periodic charging of battery 17 6 3 Replacement of battery 17 6 4 Fault finding 17 6 5 Ca...

Page 3: ...TF 200 06 0003_ENG 3 10 4 Cables 21 10 5 RF portable equipment 21 10 6 Compliance for emission and immunity standard 21 10 6 1 Emission class and group 21 10 6 2 Immunity test levels 22 11 Complaints 23 ...

Page 4: ... and handling of the product 2 General Safety This manual must be read and understood before use Always keep this manual in close proximity of the product The use installation and service of this product must be in compliance with this manual to avoid accidents and serious personal injury Never use or handle this product in other ways specified in this manual as it can result in personal safety ha...

Page 5: ...r Ropox authorized personnel Do not perform service while in use with a patient Do not allow children to operate this product unless they are under supervision from an adult that has read and understood this manual The product is not intended to play with Do not use the product if it has defects or have become damaged before being repaired or replaced The information in this manual is based on cor...

Page 6: ...oHS The product also conforms to European Medical Device Directive 93 42 EEC including amendments incorporated in Directive 2007 47 EEC Manufacturer name and address Date of production Stock number Serial number Operating interval of electrically operated equipment The use of electrical height adjustable equipment can run at maximum 2 minutes followed by an 18 minute break Type BF applied part The...

Page 7: ...17966 2016 200 kg Power supply 100 240 V 50 60 Hz I in Max 2 0 A Intermittence 2 min use 18 min pause IP rating IPX6 Ambient temperature range 5 C to 40 C Transport and storage battery included 10 C to 50 C Transport and storage without battery 5 C to 40 C Operation Relative humidity 20 to 80 non condensing Atmospheric pressure 700 hPa to 1060 hPa Materials in contact with patient Tub Cushion for ...

Page 8: ...TF 200 06 0003_ENG 8 Professional healthcare Facility Homecare healthcare Environment As defined by IEC 60601 1 2 The product is not intended for use during HF surgery ...

Page 9: ...sing acc to temp Atmospheric pressure 700 hPa to 1060 hPa Service Battery cells cannot be replaced as the battery cover cannot be closed properly afterwards For L 146 cm For L 178 cm 210 cm For L 146 cm For L 178 cm For L 210 cm For all sizes 3 2 Intended use The mobile height adjustable shower bed can be used to nurse children and adults up to 200 kg The shower bed is intended for use in private ...

Page 10: ...TF 200 06 0003_ENG 10 3 5 Dimensions 51 150 cm ...

Page 11: ... Note Remember to use locking shims All bolts are tightened with a torque of 12Nm 8 9ft lbs 4 1 2 Connection of wires in the controlbox Wires are connected according to picture and designation shown below A Motor cable 1 B Motor cable 2 C Handset D Battery accessory Note All wires are sealed with O rings make sure these a fully engaged by pushing the connectors firmly to the bottom of the socket ...

Page 12: ...will move the product downwards Warning The product should always be able to travel the full range of actuation without colliding with objects Failing to do so will compromise the product stability and introduce risk of squeezing body parts Special care should be taken in ensuring that no children or adults are sited under the product as severe injury from entrapment may occur 4 2 2 Activation and...

Page 13: ...ngaged Warning The bed guard presents a tripping hazard when exiting the product Keep your eyes on the bed guard when exiting 4 2 4 Safety in use Warning Slats may be slippery when applying soap and cause patient to slip on surface Make sure that the patient is lying down and not stand on the shower bed Make sure to deploy bed guards when using product with a patient 4 2 5 Ambient temperature Allo...

Page 14: ...e following part number 40 30762 Note The installation of the battery must be in compliance with the following installation guide 4 3 1 Mounting the battery 4 3 2 Connecting the battery Stop actuation of lifting columns Switch off the control box by disconnecting the mains plug Connect the battery to the port in the control box BA19 battery CO61 control box ...

Page 15: ...edback indicators The battery has a built in buzzer that automatically gives feedback to the user when the capacity is low This will sound when the product is operated with the hand control for adjusting the height 4 4 Electric block diagram Caution Check battery capacity before use if the battery is the only power source Make sure that the product is within distance of a power outlet to ensure co...

Page 16: ... and sterilization beyond normal cleaning may possibly change the product safety and function Chemical resistance and disinfection Detergents containing the following chemicals must not be used for cleaning or disinfection Acetone Ammonium carbonate Ethyl acetate Formic acid Hydrochloric acid Isopropanol Methyl acetate Nitric acid Phosphoric acid Pre approved cleaning agents Tradename Solution Act...

Page 17: ...ult degrading of the battery capacity Charging the battery minimum every 6 months will mitigate this 6 3 Replacement of battery Notification The battery is to be replaced after 4 years at the latest 6 4 Fault finding 6 5 Calibration of lifting columns Lifting columns are aligned and calibrated from the factory In case they need calibration to run simultaneously use the Reset procedure Problem Chec...

Page 18: ...ry Lead acid battery waste 8 Spare parts Name Part number 1 Brake wheels 40 30200 1 2 Flex Tubing Ø40 white 97001159 3 Control box CO61 97001720 4 Slats 40 25030 2 5 Motor cable 97001724 6 Control switch 97001655 7 Power cable 97000427 8 Lifting column incl bracket 40 31201 1 9 Headrest 40 25090 1 Name 40 31406 40 31410 40 31411 10 Bed guard complete 40 31196 1 40 31197 1 40 31197 1 11 Spare parts...

Page 19: ... red and green button 97003068 9 Accessories A number of configuration options are available Such options are generally ordered and delivered installed on the product at first delivery Warning Only qualified personnel may perform aftermarket installation of spare parts and accessories Caution Do not use accessories or spare parts other than those specified for the product ...

Page 20: ...or proper use and installation see 4 3 Item number 40 30762 9 2 Central locking wheels Part number Size QTY 40 30765 2 L 146 cm 1 40 30764 2 L 178 cm 1 40 30766 2 L 210 cm 1 9 3 Backrest After Pre Shipped separately Includes 2 extra slats Pre mounted at factory Part number 40 25031 Part number 40 25032 9 4 Bed guard end Part number 40 25038 ...

Page 21: ...his equipment and result in improper operation 10 5 RF portable equipment Warning Portable RF communications equipment including peripherals such as antenna cables and external antennas should be used no closer than 30 cm 12 inches to any part of the device including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result 10 6 Compliance for emi...

Page 22: ...5785 MHz 9 V m 217 Hz PM 50 duty cycle square wave Electrical fast transient burst immunity AC power ports EN 61000 4 4 2 kV 100 kHz repetition frequency Electrical fast transient burst immunity I O SIO SOP ports EN 61000 4 4 1 kV 100 kHz repetition frequency Surge immunity AC power ports EN 61000 4 5 Line to line 0 5 1 kV line to line Line to ground Not applicable the device is a Class II device ...

Page 23: ...0 06 0003_ENG 23 11 Complaints See our general terms of sale and delivery on www ropox com ROPOX A S Ringstedgade 221 DK 4700 Næstved Tel 45 55 75 05 00 Fax 45 55 75 05 50 E mail info ropox dk www ropox com ...

Reviews: