![Ronde & Schwarz NRX Getting Started Download Page 12](http://html1.mh-extra.com/html/ronde-and-schwarz/nrx/nrx_getting-started_1484988012.webp)
Preparing for Use
R&S
®
NRX
12
Getting Started 1424.7070.02 ─ 05
Risk of injury when stacking instruments
A stack of instruments can tilt over and cause injury if not stacked correctly.
Furthermore, the instruments at the bottom of the stack can be damaged
due to the load imposed by the instruments on top.
Observe the following instructions when stacking instruments:
●
Never stack more than three instruments. If you need to stack more
than three instruments, install them in a rack.
●
The overall load imposed on the lowest instrument must not exceed
500
N.
●
It is best if all instruments have the same dimensions (width and length).
If you need to stack smaller instruments on the top, the overall load
imposed on the lowest instrument must not exceed 250
N.
●
If the instruments have foldable feet, fold them in completely.
4.5
Mounting in a Rack
1. Order one of the rack adapter kits designed for the R&S NRX. For the order
number, see data sheet.
2. Follow the installation instructions provided with the adapter kit.
3.
NOTICE!
Insufficient airflow can cause overheating and thus damage the R&S
NRX.
Mounting in a Rack