Remote control commands
R&S
®
NRP18S-xx
54
User Manual 1178.3686.02 ─ 07
8.5.5
Configuring the trigger
Further information:
●
Chapter 8.5, "Controlling the measurement"
Remote commands:
.................................................................................................54
..........................................................................................54
................................................................................................55
TRIGger:EXTernal<2...2>:IMPedance
................................................................................ 56
TRIGger:ATRigger:DELay
<delay>
Effective only if
ON
.
Sets the delay between the artificial trigger event and the beginning of the actual mea-
surement
Parameters:
<delay>
Range:
0.1 to 5.0
*RST:
0.3
Default unit: Seconds
TRIGger:ATRigger:EXECuted?
Queries the number of measurements that were triggered automatically since
was set to
ON
.
In normal scalar measurements, this number can only be 0 or 1. If a buffered measure-
ment was executed, this number indicates how many results in the returned array of
measurement data were executed without a real trigger event.
Usage:
Query only
Controlling the measurement