Remote control commands
R&S
®
NRP18S-xx
59
User Manual 1178.3686.02 ─ 07
Parameters:
<source>
HOLD | IMMediate | INTernal | BUS | EXTernal | EXT1 |
EXTernal1 | EXT2 | EXTernal2
See
Chapter 8.5.2.3, "Trigger sources"
*RST:
IMMediate
TRIGger:SYNC:PORT
<sync_port>
Selects the external connection for the sync output of the power sensor. For more
information, see
.
Parameters:
<sync_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
TRIGger:SYNC:STATe
<state>
Usually used in combination with
ON
.
If enabled, blocks the external trigger bus as long as the sensor remains in the mea-
surement state. Thus, ensures that a new measurement is only started after all power
sensors have completed their measurements.
Make sure that the number of repetitions is the same for all power sensors involved in
the measurement. Otherwise, the trigger bus is blocked by any power sensor that has
completed its measurements before the others and has returned to the idle state.
Parameters:
<state>
*RST:
OFF
8.6
Configuring results
See also:
●
Chapter 8.9, "Querying measurement results"
8.6.1
Setting the power unit
The
UNIT
subsystem contains commands for setting up the power unit.
UNIT:POWer
<unit>
Sets the output unit for the measured power values.
Parameters:
<unit>
DBM | W | DBUV
*RST:
W
Example:
UNIT:POW DBM
Configuring results