![Ronde & Schwarz 1304.0004.03 User Manual Download Page 653](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004-03/1304-0004-03_user-manual_1484985653.webp)
Remote Commands
R&S
®
FPL1000
653
User Manual 1178.3370.02 ─ 09
11.6.4.7
Retrieving and Analyzing Measurement Results
The following commands retrieve and analyze measurement results for ACLR mea-
surements.
Useful commands for channel power measurements described elsewhere
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult?
●
●
CALCulate<n>:LIMit<li>:ACPower:ACHannel:RESult?
●
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:RESult?
Remote commands exclusive to channel power measurements
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult:PHZ
....................................... 653
.................................................................................. 653
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:RESult:PHZ
<State>
This command selects the way the R&S
FPL1000 returns results for power measure-
ments.
You can query results with
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:
.
Suffix:
<n>
.
<m>
<sb>
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
ON | 1
Channel power density in dBm/Hz
OFF | 0
Channel power in dBm
*RST:
0
Example:
CALC:MARK:FUNC:POW:RES:PHZ ON
Output of results referred to the channel bandwidth.
Manual operation:
See
"Channel power level and density (Power Unit)"
[SENSe:]POWer:ACHannel:MODE
<Mode>
This command selects the way the R&S
FPL1000 displays the power of adjacent chan-
nels.
Configuring and Performing Measurements