Mode d’emploi 

Moisture kit 

 "Hygromaster“ 

 Ref. 93214 

Recording Functions - PC Setup 

 
1. Connect the probes as above and confirm proper operation 
2. Disconnect the surface temperature sensor and connect the device to the serial port of the computer 
the supplied cable. 
3. Turn on the device and ensure that the side port is configured in RS232 mode (the icon 
appears) 
4. Open the software and select the 'Tools' 'Setup' (if the configuration dialog does not start 
automatically). 
5. Confirm that the serial number of the device is read correctly (not "-"): this ensures that the 
communication is established. If it does not, check the settings in 'Communication'. 
6. Click "Set time" to set the clock on the PC. 
7. Enter the required registration parameters (Recording interval / Logging interval, time to 
the first recording / Wait Before First Log, number of records and number of work / 
Number of Logs and Job ID) in the appropriate boxes and click "Set Interval" for sending 
the device. Then make sure that the recording settings are correctly configured by clicking 
"Check Interval" and check the said parameters. Finally, check that the icon         "registration" 
appears on the screen of the device. 
The configuration constructed as described above is equivalent to that described in the Configuration table 
recording. 
8. Remove the RS232 serial cable port of the device and placing the device in the environment 
control. Ensure that the Hygrostick or Humistick is connected in the upper part of the device. 
9. Press        and hold until the display shows 'OFF' (see above). 

Recording functions - Manual Recording 

 
1. It is recommended to check the clock before performing manual recording - press 
as described above to scroll through the time / date. When replacing batteries, the clock is 
to adjust from a computer as described above. 
2. The unit OFF          , press and hold 3 seconds to take a manual check point 
- This works exactly as if the recording was made with the timer as described above. 
There is no requirement to register; it is useful to configure a number of 
work to start. 
3. Similarly, the device ON, press         and hold 3 seconds to take a point 
manual recording. In this case, the port must have previously been configured mode 
surface temperature if it is to be used. 
 
 

Recording functions - Reading data 

 
1. Turn on the device and ensure that the side socket is set to RS232 mode (the icon 
appears). 
2. Open the software and select the 'Tools' 'Setup' (if the setup dialogue has not started 
automatically). 
3. Confirm that the serial number of the device is read correctly (not "-"): this ensures that the 
communication is established. If it does not, select "Get Serial Number" before checking 
settings in 'Communication'. 

4. Confirmer que le nombre des enregistrements utilisés (“Number of records used”) est celui attendu. 
Rafraîchir les données avec le bouton “Get Log Data” et fermer la fenêtre de dialogue en cliquant sur 
“OK”. 
5. Sélectionner le quatrième bouton (montre deux ordinateurs, “Dowload data”) et “All new records” dans 
la fenêtre de dialogue. Sélectionner “Latest specified number of records” pour accéder aux données plus 
anciennes si nécessaire. 
6. Quand le téléchargement est terminé, cliquer sur OK et sauvegarder. 
7. Utiliser l’outil de graphique pour un contrôle rapide (si besoin, pré sélectionner un sous-ensemble 
d’enregistrements) ou ouvrir le fichier de données sauvegardées dans un tableau pour de futures 
utilisations. 

Summary of Contents for Hygromaster

Page 1: ... Standard functions measurement of humidity and temperature 2 Tables functions off measures 3 setup 3 Functions common to configuration modes 4 Record Setup 4 Device configuration 5 Selecting the surface temperature Port 5 Recording Features PC configuration 6 recording functions manual recording 6 Recording functions reading data 6 ...

Page 2: ...the recorded measurements are displayed in a table and graph to an analysis simple and fast Hygrostick et Humistick The hygromaster is used with a detachable humidity probe for measuring relative humidity and ambient temperature Two types of sensor are available Hygrostick and Humistick The Hygrostick is optimized for measuring the humidity of 30 to 100 RH and can be used with Protimeter ankles to...

Page 3: ...the screen 4 Press repeatedly to display the room temperature RT surface temperature TS the difference in surface temperature over the dew point AT time and date the number series of the probe 5 To turn off press and hold until the display shows OFF Otherwise the unit will automatically turn off after the time set by the user 2 minutes by default after the last button press or RS232 activity Note ...

Page 4: ... probe TA TS ΔT secondary screen scrolling The time is displayed in the upper part of the screen HH MM date in the lower MM DD The serial number of the probe scrolls Storage of measurement in memory Log Press and hold 3 seconds to store measures momentary display of the number recordings and stored data b Configuration Press both buttons simultaneously see details below Button Function Display sym...

Page 5: ...es paramètres se présentent dans l ordre suivant upper part of the screen lower part of the screen Definition of PC configuration meaning Min Max Entry into the mode configuration recording the display is frozen as the buttons are pressed The logging interval is set 15 minutes example 0 No record 24H Delay before the first record 0 No delay 24H If it is not zero count the number remaining and stop...

Page 6: ...um 2 maximum 2 maximum 0 2 The new configuration is saved after pressing With the exception of the function the temperature sensor the settings are saved when the camera is off Probe function temperature of the side port automatically returns to OFF OFF when the camera is off For return to normal operating mode press or any button for 5 seconds at any configuration time f Selection of surface temp...

Page 7: ...mended to check the clock before performing manual recording press as described above to scroll through the time date When replacing batteries the clock is to adjust from a computer as described above 2 The unit OFF press and hold 3 seconds to take a manual check point This works exactly as if the recording was made with the timer as described above There is no requirement to register it is useful...

Page 8: ...g the probe to the HYGROMASTER to take the measurement After an hour turn on Raise the relative humidity value of balance HRE Installation of vinyl flooring is possible when the ERH is less than or equal to 85 to another coating if the percentage is less than 85 support is considered dry If this measure is not reached remove the probe recap the ankle with his hood which will renew the operation un...

Reviews: