PARAMETER SETTING:
WHEN SWITCHING ON THE MACHINE, THE DISPLAY SHOWS THE MIXING TIME PRESETTED:
Example: 4.5 means 4 minutes and 30 seconds / 3.0 means 3 minutes zero seconds.
To modify the mixing time ( operation possible only when mixing motor is OFF )
increase or decrease time by pushing “ 2 “ or “ 3 “ button.
The value is automatically memorized.
MIXING MOTOR SPEED
:
This value has been pre settled already by CIMO srl.
To set a different mixing motor speed ( operation possible when mixing motor is ON or OFF ):
a) push button “5” until led is flashing
b) increase or decrease motor speed by pushing “2” or “3”
c) after 5 seconds from the last pushing on 2 or 3 the setting procedure is over.
d) WHEN SWITCHING OFF THE MACHINE the parameter will go back again to the initial
CIMO value.
VIBRATION INTENSITY
:
To set the vibration intensity ( operation possible when vibration is ON or OFF ):
a) push button “7” until led is flashing
b) increase or decrease vibration intensity by pushing “2” or “3”
c) after 5 seconds from the last pushing on 2 or 3 the setting procedure is over.
d) WHEN SWITCHING OFF THE MACHINE the parameter will go back again to the initial
CIMO value.
LEAVE THE FLASK SOME MINUTES TO DRY ON THE FLASK PLATE
MEANWHILE THE OPERATOR IS WASHING AND RINSING THE BLADE AND THE
MIXING CONTAINER.
DRY ALL BEFORE A NEW OPERATION.
ALARMS ON DISPLAY
A 1 : ELECTRICAL SHORT – CONTROL BOARD DAMAGED
A 2 : MIXING MOTOR OVERHEATING
A 3 : MIXING MOTOR MECHANICAL FAULT ( too much powder loaded )
( internal blade bearing dirty )
( mixture too much dry )
A 4 : LOW TENSION AT THE UNIT
To reset the alarms: switch off and switch on the unit