The controller fitted to the scooter will let you know if a system error occurs by displaying
a flash code. These codes are as follows.
Flash
Code
Description Scooter
Condition
Action
1
Charge batteries
Scooter may
run slowly
Batteries less than 23.3v. Charge as
soon as possible.
2
Bad motor
connection.
Scooter will not
run
Check all connections between the
motor and the controller.
3
Motor short circuit.
Scooter will not
run
Check all connections between the
motor and the battery.
4
Not used.
5
Not used.
6
Inhibit circuit
activated.
Scooter will not
run
Check charger is not still connected
or a controller fault.
7
Throttle pot error.
Scooter will not
run
Check wig wag lever is in neutral
position. Faulty potentiometer.
8
Controller fault.
Scooter will not
run
Check all connections are secure.
Switch scooter on and off, this may
clear the code.
9
Magnetic brake
error.
Scooter will not
run
Check scooter is not in freewheel.
Check brake connections.
Note: Should your scooter fail to operate after checking these possible problems,
consult your local ROMA dealer immediately.
If the scooter is switched on and not operated for a period of time (approx 10 minutes) it
may enter ‘sleep’ mode. Simply switch the scooter off and on to reset the controller.