Getting Started

 

Preparation of Bioink (Adipose Tissue)

Follow the instructions below to prepare the 

adipose tissue for bioprinting.

The following process includes adipose micronization 

and washing step.

1. Adipose Tissue Micronization

Following the instructions using 4 Adipose 

Micronizers from ① to ④ in order.

A. 

Remove packaging of Adipose Micronizer and 

check for any damage.

B. 

 Attach the luer-lock syringe that contains 

the adipose tissue to the Left of the arrow (▶) 

marked on the side of the Adipose Micronizer 

① and attach an empty luer-lock syringe to the 

opposite side.

C. 

Alternately push the plunger of the syringes 

that are each attached to the opposite sides of 

the Adipose Micronizer ① so that the adipose 

tissue can pass through the Adipose Micronizer 

① at least 3~5 times in the direction of the arrow 

(▶) marked on its side.

D. 

Transfer the adipose tissue so that the 

micronized  adipose  tissue  can  finally  be 

contained in the syringe on the Right side of the 

arrow (▶) and then remove the syringe from 

Adipose Micronizer ①. 

Attach the luer-lock syringe that contains the 

adipose tissue to the Left of the arrow (▶) 

marked on the side of the Adipose Micronizer 

② and attach an empty luer-lock syringe to the 

opposite side.

E.

 Alternately push the plunger of the syringes 

that are each attached to the opposite sides of 

the Adipose Micronizer ② so that the adipose 

tissue can pass through the Adipose Micronizer 

② at least 3~5 times in the direction of the arrow 

(▶) marked on its side.

F.

 Transfer the adipose tissue so that the 

micronized adipose tissue can finally be 

CAUTION:

 

Do not use in case of damage to sterile 

                  packaging of the product components.

   

NOTE:

 

Prepare at least 4 luer-lock syringes (10~30cc)

Dr.INVIVO AI Regen KIT Manual

EUARK002_Ver.1.0.0

This product is consisted of Adipose Micronizer, 

the fat separator device and the accessories.

1. 

Adipose Micronizer ①

    Adipose Micronizer ②
    Adipose Micronizer ③
    Adipose Micronizer ④

2.

 Dispenser 1

3.

 3D Regen PCL

    3D Regen PCL Holder

4.

 Bed Film 

5.

 Dispenser 2

6. 

Dispenser 2 Cap

7.

 Dispenser 2 Tip

8.

 Dispenser 2 Tube

9.

 NFC Tag (Packaged with Dispenser 1)

1

5

7

8

9

2

3

4

Dr. INVIVO AI Regen KIT 

Components

6

Reviews: