Command Reference
R&S
®
ZNB/ZNBT
1131
User Manual 1173.9163.02 ─ 38
<PhyPt>
Test port number of the analyzer. This suffix is ignored; the gen-
erator is selected via
<Gen>
.
<Gen>
Generator number
Parameters:
<Boolean>
ON | OFF - generator is on or off
*RST:
ON
Example:
SYST:COMM:RDEV:GEN1:DEF 'Ext. Gen. 1', 'SME02',
'gpib0', '21'
Configure an R&S SME02 generator as external generator no.
1, assigning the name "Ext. Gen. 1" and a GPIB address 21.
SOUR:POW:GEN:STAT OFF
Turn the external generator off. In the "Port Configuration" dia-
log, "RF Off" is checked for generator no. 1.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:PERManent[:STATe]
<Boolean>
Defines whether the source power is permanently on.
Note
:
If
are part of the RF connection configuration, operation with
not
supported. In this case only external generators can be
used as permanent signal sources.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer
Parameters:
<Boolean>
ON - power at port
<PhyPt>
is permanently on.
OFF - power is only on for the partial measurements that require
the port as a drive port.
*RST:
OFF
Example:
SOUR:POW2:STAT ON
Switch the RF power at port no. 2 on.
SOUR:POW2:PERM ON
Activate a permanent signal source.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:REDuce[:STATe]
<ReduceAtSweepEnd>
Deprecated
; superseded by
SOURce<Ch>:POWer<PhyPt>:SWEepend:MODE
Enables/disables power reduction at sweep end.
This is a global setting.
SCPI Command Reference