Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
87
User Manual 1179.3315.02 ─ 01
OFF | 0
The analog mixer path is always used.
Example:
INP:DPAT OFF
Manual operation:
See
INPut<ip>:FILE:ZPADing
<ZeroPadding>
Enables or disables zeropadding for input from an I/Q data file that requires resam-
pling. For resampling, a number of samples are required due to filter settling. These
samples can either be taken from the provided I/Q data, or the software can add the
required number of samples (zeros) at the beginning and end of the file.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<ZeroPadding>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
INP:FILE:ZPAD ON
Manual operation:
See
INPut<ip>:FILTer:HPASs[:STATe]
<State>
Activates an additional internal high-pass filter for RF input signals from 1
GHz to
3
GHz. This filter is used to remove the harmonics of the connected instrument in order
to measure the harmonics for a DUT, for example.
This function requires an additional high-pass filter hardware option.
(Note: for RF input signals outside the specified range, the high-pass filter has no
effect. For signals with a frequency of approximately 4
GHz upwards, the harmonics
are suppressed sufficiently by the YIG-preselector, if available.)
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Configuring the Measurement