Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
114
User Manual 1179.3315.02 ─ 01
●
Remote commands exclusive to amplitude settings:
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel<ant>
................................................114
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel<ant>:OFFSet
................................... 114
...........................................................................................115
.........................................................................................115
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S
VSE or limiting the dynamic range by an
S/N ratio that is too small.
Example:
ADJ:LEV
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel<ant>
<ReferenceLevel>
This command defines the reference level (for all traces in all windows).
Suffix:
<n>
.
irrelevant
<t>
irrelevant
<ant>
(for MIMO measurements only)
Example:
DISP:TRAC:Y:RLEV -60dBm
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel<ant>:OFFSet
<Offset>
This command defines a reference level offset (for all traces in all windows).
Suffix:
<n>
.
irrelevant
<t>
irrelevant
<ant>
(for MIMO measurements only)
Parameters:
<Offset>
Range:
-200 dB to 200 dB
*RST:
0dB
Default unit: DB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Manual operation:
See
" Shifting the Display ( Offset )"
Configuring the Measurement