Remote Commands for HRP UWB Measurements
R&S
®
VSE-K149
88
User Manual 1179.3315.02 ─ 01
*RST:
0
Example:
INP:FILT:HPAS ON
Turns on the filter.
Manual operation:
See
" High Pass Filter 1 to 3 GHz "
INPut<ip>:FILTer:YIG[:STATe]
<State>
Enables or disables the YIG filter.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
Example:
INP:FILT:YIG OFF
Deactivates the YIG-preselector.
Manual operation:
See
INPut<ip>:IMPedance<ant>
<Impedance>
This command selects the nominal input impedance of the RF input. In some applica-
tions, only 50
Ω
are supported.
Suffix:
<ip>
.
1 | 2
irrelevant
<ant>
(for MIMO measurements only)
Parameters:
<Impedance>
50 | 75
*RST:
50 Ω
Default unit: OHM
Example:
INP:IMP 75
Manual operation:
See
INPut<ip>:PRESelection:SET
<Mode>
This command selects the preselector mode.
The command is available with the optional preselector.
Configuring the Measurement