![Rohde & Schwarz TSME6 User Manual Download Page 24](http://html1.mh-extra.com/html/rohde-and-schwarz/tsme6/tsme6_user-manual_1477992024.webp)
Getting Started
R&S
®
TSME6
24
User Manual 4900.0033.02 ─ 06
The following external devices are required for standard operation.
●
Connect the instrument to the power supply as described in
●
Connect the PC or notebook LAN port to the LAN port of the R&S TSME6 as
described in
Chapter 3.1.5.1, "Setting up the LAN connection to the host PC"
●
Connect the (optional) antenna's SMA-connector to the RF IN connector.
NOTE: Do not overload the input power at the RF input connector, otherwise the
input stage could be severely damaged. For maximum allowed values, see the
data sheet.
●
Connect the GPS antenna to the GPS ANT connector of the instrument for time
synchronization to a GPS signal (3 V, max. 25 mA for active antenna). To ensure
time synchronization of the R&S TSME6, it is required to have a GPS antenna con-
nected.
NOTE: A missing GPS antenna will lead over time to the point that signals cannot
be detected anymore. We recommend to connect a GPS antenna at least once per
month.
●
Connect the R&S TSME30DC / R&S TSME44DC according to the measurement
setup description in R&S TSME30DC / R&S TSME44DC Ultracompact Down-
converters manual.
Do not overload the input power at the RF input connector, otherwise the input stage
could be severely damaged. For maximum allowed values, see the data sheet.
3.1.5.3
Connecting a kensington lock
The R&S TSME6 provides a connector for a Kensington lock, which can be used to
secure a mobile device against theft. The connector is on the side panel of the instru-
ment.
Figure 3-6: Connector for a Kensington lock on theR&S TSME6