Remote control commands
R&S
®
SMM100A
713
User Manual 1179.1341.02 ─ 05
:SENSe<ch>[:POWer]:TYPE?
Queries the sensor type. The type is automatically detected.
Return values:
<Type>
string
Example:
:SENS1:TYPE?
Queries the type of sensor.
Response: NRP-Z21
The R&S NRP-Z21 sensor is used.
Usage:
Query only
Manual operation:
See
"Sensor type and serial number"
:SENSe<ch>[:POWer]:ZERO
Performs zeroing of the sensor.
Zeroing is required after warm-up, i.e. after connecting the sensor.
Note:
Switch off or disconnect the RF power source from the sensor before zeroing.
We recommend that you zero in regular intervals (at least once a day), if:
●
The temperature has varied more than about 5 °C.
●
The sensor has been replaced.
●
You want to measure very low power.
Example:
:SENS1:ZERO
Executes zeroing.
Usage:
Event
Manual operation:
See
13.15
SCONfiguration subsystem
The
SCONfiguration
subsystem contains the commands for defining the system
configuration setting, like stream mapping and used baseband configuration.
Required options
See
Example: Distributing the I/Q streams to the output connectors
The following example lists the commands necessary to fulfill this task.
// ********************************************************************************
// Map the I/Q streams to the output connectors.
// ********************************************************************************
SCONfiguration subsystem