Remote control commands
R&S
®
SMM100A
706
User Manual 1179.1341.02 ─ 05
Parameters:
<Power>
DBM | DBUV | WATT
*RST:
DBM
Example:
:SENS2:UNIT DBM
Selects dBm as unit for the measured value returned by com-
mand
READ
.
:READ2?
Response: 7.34
7.34 dBm are measured by sensor 2.
Manual operation:
See
"Level (Peak) / Level (Average)"
:SENSe<ch>[:POWer]:APERture:DEFault:STATe
<UseDefAp>
Deactivates the default aperture time of the respective sensor.
To specify a user-defined value, use the command
Parameters:
<UseDefAp>
0 | 1 | OFF | ON
*RST:
1
Example:
See
Example "Performing a power measurement with a fixed fil-
Manual operation:
See
:SENSe<ch>[:POWer]:APERture:TIMe
<ApTime>
Defines the aperture time (size of the acquisition interval) for the corresponding sensor.
Parameters:
<ApTime>
float
Range:
depends on connected power sensor
Increment: 1E-9
*RST:
depends on connected power sensor
Example:
See
Example "Performing a power measurement with a fixed fil-
Manual operation:
See
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect
<Select>
Several S-parameter tables can be stored in a sensor. The command selects a loaded
data set for S-parameter correction for the corresponding sensor.
Parameters:
<Select>
float
*RST:
0
SENSe, READ, INITiate and SLISt subsystems