Network operation and remote control
R&S
®
SMM100A
581
User Manual 1179.1341.02 ─ 05
Remote device
VNC client
Requirements
Characteristics
●
Web browser with HTML5
Web sockets
must be supported.
Slower than the other modes.
No additional installation or acti-
vation required.
No security concern.
Smart device
(Tablet/ smartphone)
●
Dedicated client App
App
must be installed.
Fast, supports several options like
full screen mode or auto-login.
●
Web browser with HTML5
Web sockets
must be supported.
Support of QR code scanning
Slower than a dedicated App.
Remote file access (FTP, SAMBA/SMB)
●
A remote client accesses the instrument's file system, using the protocols FTP (file
transfer protocol) and SAMBA/SMB (server message block).
●
The protocols enable you to transfer files from or to the instrument and to get direct
access to its file sharing directory
share
.
How to:
Chapter 10.9, "How to transfer files from and to the instrument"
"Activating SMB version 1.0/2.0 client and server"
12.2
Remote control interfaces and protocols
The instrument supports various interfaces for remote control.
overview on the connectivity:
Table 12-2: Remote control interfaces and protocols
Interface
Protocols, VISA
*)
address string and library
Remarks
Local area network
(LAN)
●
HiSLIP
High-Speed LAN Instrument Protocol
(IVI-6.1)
TCPIP::host address::hislip0[::INSTR]
VISA
●
VXI-11
TCPIP::host address[::
LAN device name][::INSTR]
VISA
●
Socket communication
(Raw Ethernet, simple
Telnet)
TCPIP::host address[::
LAN device name]::<port>::SOCKET
VISA or socket controller
The LAN connector is at the rear panel of the
instrument.
The interface is based on TCP/IP, see
details on the address information.
USB
●
USBTMC
USB::<vendor ID>::<product ID>::
<serial number>[::INSTR]
VISA
The USB Device connector is at the rear panel
of the instrument.
For a description of the interface, see
Remote control interfaces and protocols