Multi-instrument setups
R&S
®
SMM100A
460
User Manual 1179.1341.02 ─ 05
Connector
Direction/Color
Enabled in the "I/Q Stream Mapping"
RF
-
Yes
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:DIRection?
:SCONfiguration:EXTernal:IQOutput<ch>:DIRection?
:SCONfiguration:EXTernal:RF<ch>:DIRection?
External Instrument
Accesses a dialog with settings that help you to establish the connection to the exter-
nal instruments, see
Chapter 9.4.2.2, "External instrument configuration settings"
Note:
Each external instrument must be connected via LAN or USB to the R&S
SMM100A.
In most test setups, all instruments must use a common reference frequency signal,
either from an external common reference source or using the distributed reference
frequency of the R&S
SMM100A.
I/Q Connection
Depends on the interface type.
●
Analog I/Q interfaces:
Selects the type of the I/Q output signal
●
Digital I/Q interfaces:
Indicates the status of the I/Q connection of the digital interfaces.
The indication includes a status LED and an icon, where the icon depends on the
connector (Dig I/Q or HS Dig I/Q) used.
Indication
Designation
green LED and an icon
The Dig I/Q/HS Dig I/Q interface is used as an
input
connector
yellow LED and an icon The Dig I/Q/HS Dig I/Q interface is used as an
output
connector
icon without an LED
An external instrument is connected to the Dig I/Q interface, but no direction is
specified
No icon displayed
No connection
"Single Ended"/"Differen-
tial"
Selects the type of the I/Q output signal at the I/Q output connector
Works like the parameter
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:IQConnection:STATe?