Configuring the RF signal
R&S
®
SMM100A
398
User Manual 1179.1341.02 ─ 05
Remote command:
[:SOURce<hw>]:CORRection[:STATe]
User Correction
Indicates the corrected level value for a specific frequency point.
Remote command:
[:SOURce<hw>]:CORRection:VALue?
UCOR Data
Accesses the standard "File Select" function of the instrument. The provided navigation
possibilities in the dialog are self-explanatory.
Files with user correction values are files with predefined file extension
*.uco
. When a
file is selected, the dialog indicates the filename.
You can create the file internally in the table editor or externally.
●
To select an existing file, select "Select List > navigate to the file *.uco > Select"
●
Use the general editor function to create internally new file or to edit an existing
one.
●
Use the standard file manager function to load externally created files to the instru-
ment.
Remote command:
[:SOURce]:CORRection:CSET:CATalog?
[:SOURce<hw>]:CORRection:CSET[:SELect]
[:SOURce]:CORRection:CSET:DELete
Edit UCOR Data
Opens the build-in table editor to define a new correction table or edit an existing one.
See also:
●
●
●
Chapter 7.12.3.3, "Fill with sensor"
7.12.3.2
List editor
The "User Correction" and "List Mode" dialogs provide a build-in list editor for defining
the corresponding value pairs.
The table and navigation functions of the UCOR and list mode editors have the same
structure and are therefore summarized below. The access is specified for both modes,
the fields and functions are explained using the example of the list mode.
Access to "Edit List Mode Data":
► "RF" > "Sweep/List" > "List Mode" > "List Mode Data" > "Edit List Mode Data"
The editor for list mode provides a table with RF frequency and power values, an
extra column for defining variable dwell times, and standard navigation functions.
Improving level performance