Maximal possible deviation is selected automatically, if one of the following applies:
●
The selected deviation is too high for the particular frequency
●
The selected frequency is outside of the range where deviation is possible.
A warning message indicates this situation, too.
Remote command:
Deviation Mode ← PhiM settings
Enables you to couple the LF signal of both paths and determines the mode for fixing
the total deviation.
The deviation sum 2 rad at a maximum. When the sum is exceeded, the
R&S
SMM100A adjusts the settings automatically.
"Uncoupled"
Enables you to adjust the deviation for each path independently.
"Fixed Total"
Couples the deviation of both paths. If you change the deviation of
any path, the R&S
SMM100A automatically adjusts the value of the
other path. The sum always results in the set
.
"Fixed Ratio"
Couples the deviation ratio of both paths. If you change the deviation
of any path, the R&S
SMM100A adjusts the value of the other path.
Remote command:
[:SOURce<hw>]:PM:DEViation:MODE
Total Deviation ← PhiM settings
In
, sets the sum of the signal deviation for path 1 and
path 2.
of both paths always sum up to the value of the total deviation.
Behavior when changing any depth settings:
●
Changing the total deviation adjusts the deviations of both paths proportionally.
●
Changing the deviation of one path adjusts the deviation of the other path. The
sum always results in the set total deviation.
Remote command:
[:SOURce<hw>]:PM:DEViation:SUM
Ratio Path2/Path1 ← PhiM settings
Sets the deviation ratio (path2 to path1) in per cent.
Example:
If the deviation in path1 is 10 rad and the ratio is 50%, the deviation in path 2 is auto-
matically set to 5 rad.
Remote command:
Mode ← PhiM settings
Selects the mode of the phase modulation.