Remote control commands
R&S
®
SMM100A
760
User Manual 1179.1341.02 ─ 05
NOISe
Uses the internally generated noise signal.
INTernal
Uses the internally generated signal of LF1.
EXTernal
Uses an external LF signal (EXT1).
*RST:
LF1 <PM1>; LF2 <PM2>
Example:
See
Example "Performing phase modulation"
Manual operation:
See
[:SOURce<hw>]:PM:MODE
<Mode>
Selects the mode for the phase modulation.
Parameters:
<Mode>
HBANdwidth | HDEViation | LNOise
HBANdwidth
Sets the maximum available bandwidth.
HDEViation
Sets the maximum range for
Φ
M deviation.
LNOise
Selects a phase modulation mode with phase noise and spuri-
ous characteristics close to CW mode.
*RST:
HBANdwidth
Example:
See
Example "Performing phase modulation"
Manual operation:
See
[:SOURce<hw>]:PM:DEViation:MODE
<PmDevMode>
Selects the coupling mode. The coupling mode parameter also determines the mode
for fixing the total deviation.
Parameters:
<PmDevMode>
UNCoupled | TOTal | RATio
UNCoupled
Does not couple the LF signals.
The deviation values of both paths are independent.
TOTal
Couples the deviation of both paths.
RATio
Couples the deviation ratio of both paths
*RST:
UNCoupled
Example:
See
Example "Using combined LF signals"
Manual operation:
See
SOURce subsystem