Configuring the RF signal
R&S
®
SMM100A
315
User Manual 1179.1341.02 ─ 05
2. Select "Reference Output".
In the "Reference Output" tab, you can set the reference frequency value at the
output connectors.
As a result of parameter dependencies, "Preset This Parameter" sometimes does
not affect output dialogs.
The remote commands required to define these settings are described in
ter 13.16.17, "SOURce:ROSCillator subsystem"
Settings:
Reference Output/1 GHz Reference Output
............................................................... 315
Reference Output/1 GHz Reference Output
Selects frequency reference output signal for downstream instruments.
"Off"
Deactivates the reference signal output.
"10
MHz"
Derives a signal with 10
MHz frequency from the internal reference
oscillator and provides this signal at the output.
"100
MHz"
Option: R&S
SMM-K703
Provides the 100
MHz reference frequency signal at the output.
"1
GHz"
Option: R&S
SMM-K703
Provides the 1
GHz reference frequency signal at the output.
"Input Signal (loop through)"
Option: R&S
SMM-K703/-K704
Passes the external reference frequency to the output directly.
Remote command:
[:SOURce]:ROSCillator:OUTPut:FREQuency:MODE
[:SOURce]:ROSCillator:OUTPut:ALTernate:FREQuency:MODE
Show Connector
Accesses a dialog that displays the physical location of the selected connector on the
front/rear panel of the instrument.
A blinking LED on the front/rear panel indicates the selected connector, too (see
ter 11.2.2.4, "Status indicators"
Reference oscillator