Maintenance, storage and disposal
R&S
®
SMM100A
1023
User Manual 1179.1341.02 ─ 05
Remote command:
Temperature Offset
Displays the temperature difference, comparing the temperature of the last adjustment
to the current instrument temperature.
A green checkmark indícates that the offset is within the permitted range. If the temper-
ature deviates more than
±
5
K, the instrument indicates a warning icon.
Remote command:
Information
Displays information to the current adjustment state.
Remote command:
:CALibration<hw>:ALL:INFormation?
Adjust I/Q Modulator Current Frequency
Starts the adjustment for the I/Q modulator for the currently set frequency and base-
band gain. The I/Q modulator is adjusted regarding carrier leakage, I/Q imbalance and
quadrature.
See also
"To trigger the internal correction routine of the I/Q modulator"
Remote command:
:CALibration<hw>:IQModulator:LOCal?
Adjust External Frequency Converter
Calibrates the parameters of an externally connected R&S SZU IQ Upconverter.
The function is available when an R&S SZU is connected and activated.
Remote command:
:CALibration<hw>:FREQuency:CONVerter:EXTernal?
Continue Adjustment on Error
Continues the calibration even though an error was detected. By default adjustments
are aborted on error.
This function is password-protected. Unlock the protection level 1 to access it.
Remote command:
:CALibration<hw>:CONTinueonerror
Log Debug Info
Activates logging of the internal adjustments.
This function is password-protected. Unlock the protection level 1 to access it.
Remote command:
Performing maintenance tasks