Test sequence
R&S®SML / R&S®SMV03
1090.3123.12 10.8
E-6
Reference frequency
Internal reference output
Note
:
Let the R&S SML / R&S SMV03 warm up for at least 2 hours before starting a measurement.
Test setup
Connect test receiver (spectrum analyzer mode) to REF EXT output
on rear panel of R&S SML / R&S SMV03.
Settings on the test receiver - Center frequency = 10 MHz
- Span = 20 Hz
- Resolution bandwidth = 10 Hz
Measurement
Set marker on peak and read off frequency.
Analysis
The frequency deviation must not exceed the sum of the deviations
resulting from the frequency error in the rated temperature range and
from aging.
External reference input
Test setup
Connect test receiver (spectrum analyzer mode) to RF connector on
R&S SML / R&S SMV03.
Connect signal generator to REF IN connector on rear panel of R&S
SML / R&S SMV03.
Connect references of test receiver and signal generator
Settings on the test receiver - Center frequency = test frequency of R&S SML / R&S SMV03
- Span = 100 Hz
- Resolution bandwidth = 10 Hz
Settings on R&S SML /
R&S SMV03
- Switch reference oscillator to external
Settings on the signal
generator
- Frequency = 10 MHz
- Level = 7 dBm
Measurement
Set marker on peak and read off frequency
Analysis
The frequency deviation must not exceed the sum of the deviations
resulting from the frequency error in the rated temperature range and
from aging.
Summary of Contents for SML01
Page 4: ......
Page 14: ......
Page 22: ......
Page 32: ......
Page 34: ......
Page 36: ......
Page 38: ......
Page 40: ......
Page 42: ......
Page 44: ......
Page 56: ......
Page 62: ......
Page 77: ...R S SML R S SMV03 Menu Summary for R S SML 1090 3123 12 E 6 3 15 Menu Summary for R S SML ...
Page 78: ...Menu Summary for R S SMV03 R S SML R S SMV03 1090 3123 12 E 6 3 16 Menu Summary for R S SMV03 ...
Page 148: ......
Page 240: ......
Page 242: ......