Remote Control Commands
R&S
®
SMF100A
545
Operating Manual 1167.2319.02 ─ 12
[:SOURce<hw>]:PULM:TRAin:SELect
<Filename>
Selects the specified pulse train file. If a new file is to be created, the name can be
entered here. The file is created if it does not yet exist. The file selected here is availa-
ble for the further processing steps (editing) and is used in the instrument when the
pulse train mode is activated.
The files are stored with the fixed file extensions
*.pulstrn
in a directory of the
user's choice. The directory applicable to the command is defined with the command
MMEMory:CDIR
.
*RST
does not affect data lists.
Parameters:
<Filename>
string
Example:
MMEM:CDIR '/var/user/Lists'
selects the directory for the pulse train files.
PULM:TRA:SEL ‘P_INCR'
selects
P_INCR
for editing.
P_INCR
is created if it does not yet
exist.
Options:
R&S SMF-K27 (Pulse Train)
Manual operation:
See
" Pulse Train Data – Pulse Generator "
[:SOURce<hw>]:PULM:TRIGger:EXTernal:GATE:POLarity
<Polarity>
Selects the polarity of the Gate signal.
The signal is supplied via the PULSE IN connector.
Parameters:
<Polarity>
NORMal | INVerted
*RST:
NORMal
Example:
PULM:TRIG:EXT:GATE:POL NORM
The pulse signal is generated while the gate signal is high.
Manual operation:
See
" Gate Input Polarity - Pulse Generator "
[:SOURce<hw>]:PULM:TRIGger:EXTernal:IMPedance
<Impedance>
Selects the impedance for external pulse trigger.
Parameters:
<Impedance>
G50 | G10K
*RST:
G50
Example:
SOUR:PULM:TRIG:EXT:IMP G50
selects 50 Ohm as the trigger impedance for the external pulse
trigger.
Manual operation:
See
SOURce Subsystem